7-2 Description of the Functions

Table 5 Functions

No.

Function

SYSTOLIC

DIASTOLIC

Description

 

 

 

 

 

 

display

display

 

F01

 

 

 

 

 

f01

off

Not available.

 

 

 

 

 

F02

Printer setting

f02

off

No printing

 

(See “7-3

 

1

High-speed printing

 

Printing

 

 

 

 

 

2

3-line printing

 

Samples” for

 

 

what each

 

3

3-line printing

 

method printing

 

 

 

 

 

4

Graph printing

 

looks like.)

 

 

TM-2655P only

 

5

Table printing

F03

 

 

 

 

 

f03

off

No bit pattern printing

 

 

 

 

 

 

 

on

Bit pattern printing

F04

Displaying time

f04

999

The results remain in the display unless

 

 

 

 

 

 

 

 

another operation is performed.

 

 

 

 

 

 

 

5

Displays the results for 5 seconds.

 

 

 

 

 

 

 

10

Displays the results for 10 seconds.

 

 

 

 

 

 

 

20

Displays the results for 20 seconds.

F05

External

Ff05

1

Terminals 1 and 2 to a PC

 

connection

 

2

Terminal 1 to a PC

 

 

 

 

 

 

 

 

Terminal 2 to a card reader

 

 

 

 

 

 

 

3

Terminal 1 to a scale

 

 

 

 

 

 

 

 

Terminal 2 to a PC

F06

Pressurization

Ff06

au

Automatic pressurization

 

 

 

 

 

 

 

160

Pressurizes the cuff up to 160 mmHg.

 

 

 

 

 

 

 

180

Pressurizes the cuff up to 180 mmHg.

 

 

 

 

 

 

 

200

Pressurizes the cuff up to 200 mmHg.

F07

Channel 1 baud

Ff07

120

1200 bps

 

rate

 

240

2400 bps

 

 

 

 

 

 

 

 

 

 

 

 

 

 

480

4800 bps

 

 

 

 

 

 

 

960

9600 bps

F08

Channel 2 baud

Ff08

120

1200 bps

 

rate

 

240

2400 bps

 

 

 

 

 

 

 

 

 

 

 

 

 

 

480

4800 bps

 

 

 

 

 

 

 

960

9600 bps

F09

 

 

 

 

 

Ff09

H9

Not available.

 

 

 

 

 

F10

Human sensor

Ff10

off

Does not detect a human presence.

 

 

 

 

 

 

 

on

Detects a human presence.

F11

Automatic

Ff11

off

Does not output the results

 

 

 

 

 

 

 

 

automatically.

15

Image 23
Contents Automatic Blood Pressure Monitor Page Important information that helps users operate the device Compliance with the Australian EMC Frame Work ComplianceGrounding Safety PrecautionsWhen installing and storing the device General PrecautionsMust be disposed of according to the applicable regulations Page Contents Troubleshooting External Dimensions Introduction FeaturesItems contained Unpacking and InspectionPerformance specifications SpecificationsMain Unit Part NamesRear Panel SymbolsProper installation InstallationBe sure to insert your arm to the shoulder OperationMeasurement results Setting Example Setting the ClockClock Adjustment Mode Page Paper installation procedure Installing the Printer PaperFunction Setting Mode Setting the FunctionsF02 Off Description of the FunctionsLine printing Table printing Bit pattern printing Printing SamplesChannel 1 Miniature DIN 8-pin Communication SpecificationsChannel 2 D-sub 9-pin Cable connection diagramReplacing the Arm Cuff Cover MaintenanceChecking the Counter Error CodesReplacing the Fuses Cleaning MaintenanceOptions and Consumables Troubleshooting External dimensions External Dimensions