A&D TM-2655P manual Channel 2 D-sub 9-pin, Cable connection diagram

Page 26

Fig.15 Cable connection diagram

8-2 Channel 2: D-sub 9-pin

CAUTION

zThe personal computer and medical equipment connected to the device are not allowed to be in the patient area.

zThe personal computer used must conform to IEC60950

Specifications

Standard

Transmission

Baud rate

Data bits

Parity bit

Stop bits

Code

Table 7 Channel 2 specifications

Conforms to EIA RS-232C Asynchronous, half-duplex

1200, 2400, 4800, 9600 bps (Can be changed in the function setting mode “F08”.)

8bits None

1bit

ASCII

Equipment available for connection

Channel 1: Personal computer

Channel 2: Card reader

Pin assignment

Pin No.

Signal name

Direction

Description

 

1

-

-

-

 

2

RXD

In

Receive data

 

3

TXD

Out

Send data

 

4

DTR

Out

Data terminal ready

 

5

GND

-

Signal ground

Note

6

DSR

In

Data set ready

The protocol depends

7

RTS

Out

Ready to send

8

CTS

In

Clear to send

on the equipment

9

-

-

-

connected.

Fig.16 Pin assignment

Cable connection between the device and a personal computer

TM-2655/TM-2655P

 

 

 

Personal computer

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

18

Image 26
Contents Automatic Blood Pressure Monitor Page Important information that helps users operate the device Compliance Compliance with the Australian EMC Frame WorkSafety Precautions GroundingGeneral Precautions When installing and storing the deviceMust be disposed of according to the applicable regulations Page Contents Troubleshooting External Dimensions Features IntroductionUnpacking and Inspection Items containedSpecifications Performance specificationsPart Names Main UnitSymbols Rear PanelInstallation Proper installationOperation Be sure to insert your arm to the shoulderMeasurement results Setting Example Setting the ClockClock Adjustment Mode Page Installing the Printer Paper Paper installation procedureSetting the Functions Function Setting ModeDescription of the Functions F02 OffPrinting Samples Line printing Table printing Bit pattern printingCommunication Specifications Channel 1 Miniature DIN 8-pinCable connection diagram Channel 2 D-sub 9-pinMaintenance Replacing the Arm Cuff CoverChecking the Counter Error CodesReplacing the Fuses Maintenance CleaningOptions and Consumables Troubleshooting External Dimensions External dimensions