A&D TM-2655P manual Troubleshooting

Page 31

10. TROUBLESHOOTING

If the TM-2655/TM-2655P does not function properly or an error code appears, try the following corrective actions.

Table 10 Troubleshooting

Problem

 

Check this

Corrective action

Nothing appears in

Is the power cable connected

Connect the power cable properly.

properly?

the display, even

 

 

 

 

 

 

Replace both fuses with new ones.

when the power is

 

 

 

 

 

Has a fuse burned out?

(Size 5x20mm. Rated voltage and current

turned on.

 

 

 

 

 

are specified on the rear panel).

 

 

 

 

 

 

The cuff will not

Is the arm cuff cover installed

Refer to “9-1 Replacing the Arm Cuff

inflate.

with high tension (too tight)?

Cover” to re-install the cover.

 

Is your posture proper?

Place your arm at your heart level, relax

 

and remain still during the measurement.

 

 

 

 

 

 

 

Do you remain still?

Do not move your arm during the

 

measurement. Relax and remain still.

 

 

 

 

 

 

The device will not

 

 

 

 

 

Wearing a thick jacket may cause a faint

measure.

 

 

 

 

 

pulse, and result in a measurement error.

(Eerr appears.)

 

 

 

 

 

Take off the jacket.

 

 

 

 

 

 

This device bases its measurements on

 

 

 

 

 

 

the heartbeat. If you have a very weak or

 

 

 

 

 

 

irregular heartbeat, the device may have

 

 

 

 

 

 

difficulty determining your blood pressure.

 

Is the printer paper installed?

Refer to “6. INSTALLING THE PRINTER

 

(pe appears.)

PAPER” to install the printer paper.

 

Is the printer head up?

Lower the lever and lock the printer paper

 

(HU appears.)

in position

The printer does

Is the printer paper inserted

Refer to “6. INSTALLING THE PRINTER

not print.

straight or is the printer paper

PAPER” to re-install the printer paper.

 

jammed?

 

 

 

 

 

 

 

 

Refer to “7-2 Description of the

 

Is the printer setting correct?

Functions” to select the correct printer

 

 

 

 

 

 

setting.

Note

If the actions described above do not solve the problem, contact the dealer. Do not attempt to repair the device yourself.

23

Image 31
Contents Automatic Blood Pressure Monitor Page Important information that helps users operate the device Compliance with the Australian EMC Frame Work ComplianceGrounding Safety PrecautionsWhen installing and storing the device General PrecautionsMust be disposed of according to the applicable regulations Page Contents Troubleshooting External Dimensions Introduction FeaturesItems contained Unpacking and InspectionPerformance specifications SpecificationsMain Unit Part NamesRear Panel SymbolsProper installation InstallationBe sure to insert your arm to the shoulder OperationMeasurement results Clock Adjustment Mode Setting the ClockSetting Example Page Paper installation procedure Installing the Printer PaperFunction Setting Mode Setting the FunctionsF02 Off Description of the FunctionsLine printing Table printing Bit pattern printing Printing SamplesChannel 1 Miniature DIN 8-pin Communication SpecificationsChannel 2 D-sub 9-pin Cable connection diagramReplacing the Arm Cuff Cover MaintenanceReplacing the Fuses Error CodesChecking the Counter Cleaning MaintenanceOptions and Consumables Troubleshooting External dimensions External Dimensions