Understanding VSI Controller Diagnostics Codes (cont.)
3.5 Five Bars - Right Motor Wiring Trip
The right hand motor has a short circuit to a battery connection.
Test Right Motor Short
Check that the batteries are fully charged and in good con- dition, and check all cables and connections. Check the connections to the right motor, look for a loose or damaged connector. Take a resistance reading from pin 2 to pin 9 see (figure A3.5.1). Take a resistance reading from pin 2 to pin 7 or pin 8 see (figure A3.5.2), if the all the circuits are open (resistance is greater than 10K ohms), then replace the controller.
A3.5.1
A3.5.2
![]()
If the all the circuits are open (resistance is greater than 10K ohms), then replace the controller.
If the reading is short (resistance is less than 10 K ohms) on any of the readings, proceed to check the
A3.5.3
Test 1 |
| Test 2 |
| ||
|
|
|
Test 3 |
| Test 4 | ||
| ||||
|
|
|
|
|
![]()
If all of the readings are open, then replace the main harness. ![]()
If any of the readings are short, then replace the right motor.
SUNRISE MEDICAL | ASPIRE SERVICE MANUAL | 2003 | PAGE 8 |