A&D GH-120/GH-200/GH-300/GH-202/GH-252 manual Details of the Function Table

Page 36

10.2.Details of the Function Table

Class

Item and Parameter

 

 

Description

 

Cond

 

0

Fast response, sensitive value

 

FAST

 

Can be changed by response

 

1

 

 

 

 

Condition

 

 

 

 

MID.

 

adjustment.

 

2

Slow response, stable value

 

 

 

 

SLOW

 

 

 

 

 

 

 

 

5t-b

 

0

Stable range is ±1 digit

 

 

 

 

The stabilization indicator illuminates

 

 

1

 

#1

 

when the display fluctuation is within

 

Stability band width

 

 

 

 

 

2

Stable range is ±3 digits

 

 

 

 

the range.

 

 

 

 

 

 

 

 

 

 

ba5fnc

 

 

0

OFF

 

 

 

 

 

 

 

trc

 

1

Normal

 

 

 

 

Keeps zero display by tracking zero

Environment

 

 

 

 

 

Zero tracking

 

2

Strong

 

 

 

 

drift.

 

 

 

 

 

Display

 

 

3

Very strong

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

5pd

 

0

5 times/second

 

 

 

 

Period to refresh the display

 

Display refresh rate

 

1

10 times/second

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pnt

 

0

Point (.)

 

 

 

 

Decimal point format

 

Decimal point

 

1

Comma (,)

 

 

 

 

 

 

 

 

 

 

 

 

 

 

p-on

 

0

OFF

 

 

 

 

Turns on the weighing mode display

 

Auto display-ON

 

1

ON

 

 

 

 

when the AC adapter is connected.

Cl adj

Refer to "10.7. Clock and Calendar Function"

 

 

 

 

The time and date are added to the

Clock

 

 

 

 

output data.

 

 

 

 

 

 

 

 

 

 

 

0

Key mode

 

 

 

 

Accepts the

PRINT

key only when

 

 

 

 

 

 

 

the display is stable.

 

 

 

 

 

 

 

 

 

 

 

 

1

Auto print mode A

 

 

 

 

Outputs data when the display is

 

prt

 

 

(Reference = zero)

 

 

 

 

 

 

 

 

 

 

 

stable and conditions of ap-p, ap-b

 

Data output mode

 

 

Auto print mode B

 

 

 

 

 

 

2

 

 

 

 

and the reference value are met.

 

 

 

(Reference = last stable value)

 

With data 0, outputs data

 

 

 

 

Stream mode /

 

 

 

 

 

 

 

3

 

 

 

 

continuously; with data 2, uses

dout

 

 

 

Interval memory mode

 

 

 

 

interval memory.

 

 

0

Plus only

 

 

 

 

Displayed value>Reference

Data output

ap-p

 

 

 

 

 

 

1

Minus only

 

 

 

 

Displayed value<Reference

 

Auto print polarity

 

 

 

 

 

 

 

2

Both

 

 

 

 

Regardless of displayed value

 

 

 

 

 

 

 

 

ap-b

 

0

10 digits

 

 

 

 

Difference between reference value

 

 

1

100 digits

 

 

 

 

and displayed value

 

Auto print difference

 

 

 

 

 

 

 

2

1000 digits

 

 

 

#2

 

 

 

 

 

 

 

 

 

 

 

 

 

data

 

0

Not used

 

 

 

 

Related items:

 

 

1

Stores weighing data

 

 

 

 

 

Data memory

 

 

 

 

 

prt,int, d-no, 5-td, info

 

 

2

Stores calibration data

 

 

 

 

 

 

 

 

 

 

 

 

 

 

: Factory settings.

#1 The unit of minimum display is digit.

Example: If 1 mg display is selected using the RANGE key for the GH-300, 1 mg is one digit.

#2 Usable minimum display of the balance is one digit.

Example: In gram display, one digit is 0.01 mg for the GH-252 and 0.1 mg for the GH-300.

34

Image 36
Contents GH-120 GH-200 GH-300 GH-202 GH-252 1WMPD4000891CPage Contents Underhook Density Measurement Interface And CommunicationIntroduction About This ManualFeatures ComplianceCompliance With FCC Rules Compliance With EMC DirectivesPage Unpacking the Balance Precautions Installing the BalanceBefore Use During Use ShockPower Supply After UseDisplay Symbols and Key Operation Key When pressed and released When pressed and heldDisplay Key operationWeighing Units Units101 00001 Counting mode Pcs Percent mode Densitymode Press the Print Changing the UnitsExamples Unit Display Gram Balance displaysUnit setting example Precautions for the weighing operation Basic Operation Gram ModeRead . Display symbols and Key operation before operation WeighingDual Range Weighing range Available minimum display01mg 1mgCounting Mode PCS Storing a sample unit mass Weighing input modeCounting operation Counting mode using the Acai function Mark turns off after Unit mass is improved RepeatPercent Mode % Selecting the percent modeStoring the 100% reference mass Reading the percentageResponse Adjustment Automatic Response AdjustmentAdvise Manual Response Adjustment Key untilDisplayed, then release the key Press Key again quicklyCalibration Calibration GroupAutomatic self calibration due to changes in temperature Automatic Self CalibrationOne-Touch Calibration MarkCalibration Test Using the Internal Mass Press and hold the CAL Key until is CCCalibration Using an External Weight Calibration Test Using an External Weight Key................. The key To cancel the operationKey ..............The key to store the new mass value Even if the AC adapter is removed,Correcting the Internal Mass Value Function Switch and Initialization Permit or InhibitPress the Print key To cancel this operation, press Initializing the BalanceKey to display Setting the Function Table Display symbol and keys Function TableSetting procedure Function table is displayed PressKey to select a class Press Key to enter the class Key to select a item PressSetting example Details of the Function Table Factory settings IntData output Time/Date output ID number output5if Timeout AK, Error codeErCd Interface Ct5Description of the Class Environment, Display Description of the Item Data Output Mode Key modeStream mode Auto print modes a and BInterval memory mode Description of the Item Data FormatStandard format DP Dump print formatKF format MT formatNU numerical format Data Unit Terminator Stable value Unstable valueCSV format Data numberDate TimePositive error Negative error SpaceData Format Examples Stable Unstable OverloadUnits Setting the time Clock and Calendar FunctionOperation Confirming the timeSetting the date Quitting the operationSet the date Store Setting the ID Number ID Number and GLP ReportCalibration report using the internal mass GLP ReportKey operation Calibration test report using an internal mass Balance returns to weighing mode after this testCommad Calibration report using an external weight Calibration test report using an external weight When using preset target value, proceed to stepTitle block and end block Title blockWeighing data Returns to the weighing mode Data MemoryReleasing Clr or err Press and hold the Print key until the keyData Memory for Weighing Data Storing the weighing dataMemory full Interval memorySetting the function table Parameter settings for each output mode are as followsAdditional parameter settings, as follows Enabling the data memory functionRecalling the memory data Transmitting all memory data at one timeDeleting all memory data at one time Data Memory for Calibration and Calibration Test Storing the calibration and calibration test dataTransmitting the memory data Deleting data stored in memoryUnderhook Density Measurement Formula to obtain the densityPrior to measurement Changing the function table Density of a liquidRelation between the water temperature and density Entering the density of a liquidMeasuring the density Output or store Density display Turns off. lightsPress the modes Key to proceed to otherStandard Input and Output Interface RS-232C InterfaceSub 25 pin assignments Connection to Peripheral Equipment Connection to an AD-8121B printerUsing Windows Communication Tools Software WinCT Using the WinCT software, the balance can do the followingConnection to a computer and the use of WinCT RsComCommands Command ListAcknowledge Code and Error Codes Control Using CTS and RTSSettings Related to RS-232C Command Examples On command turning on the balanceCommand Re-zeroing the display CAL command Calibration with internal mass Error codeCommand ?PT command Net value Tare value Weighing with a tareCommand Execution PT command to set a target weight of 10 g Zero displayZero Maintenance Error CodesTreatment of the Balance Sample mass error Unit mass errorAutomatic response adjustment zero error Automatic response adjustment unstable errorOther Display Checking the Balance Performance and Environment Checking that the balance performs properlyAsking for Repair Specifications External Dimensions Unit mmOptions and Peripheral Equipment GH-02 USB interfaceGH-08 Ethernet Interface AD-8121B PrinterAD-1653 Density Determination Kit AD-1683 DC Static EliminatorAX-USB-9P-EX USB Converter 9970 g/cm3AX-SW128 Foot switch AX-T-314A-S External key PlugOther options Terms/Index TermsIndex Options 5pd Out5tart Tare PU5eMemo Memo