10.4.Description of the Item "Data Output Mode"

The parameter setting of "Data output mode (prt)" applies to the performance when the "Data memory (data)" parameter is set to "2" (to store the weighing data) and when the data is transmitted using the RS-232C interface.

Key mode

When the

PRINT

key is pressed with the stabilization indictor turned on, the balance

outputs or stores the weighing data and the display blinks one time.

Required setting

dout prt 0 Key mode

Auto print modes A and B

When the displayed value is stable and the conditions of "Auto print polarity", "Auto print difference" and reference value are met, the balance outputs or stores the weighing data. When the PRINT key is pressed with the stabilization indictor turned on, the balance outputs or stores the data and the display blinks one time.

Auto print modes A

Example

For weighing each time a sample is placed and removed, with "ar-d"

 

set to "1" (to adjust zero after the data is output).

Required setting

dout

prt 1

Auto print mode A (reference = zero)

 

dout

ap-p

Auto print polarity

 

dout

ap-b

Auto print difference

 

dout

ar-d 1

Zero after output

Auto print modes B

Example

For weighing while a sample is added.

Required setting

dout

prt 2

Auto print mode B (reference = last stable value)

 

dout

ap-p

Auto print polarity

 

dout

ap-b

Auto print difference

Stream mode

The balance outputs the weighing data continuously regardless of the display condition. When the display refresh rate is set to 5 times / second (5pd 0), the data output rate is also set to the same 5 times / second. The display does not blink in this mode. The interval memory mode is used when the "Data memory (data)" parameter is set to "1"

(to store the weighing data).

 

 

Example

For monitoring data on a computer.

Required setting

dout

prt 3

Stream mode

 

dout

data 0

Data memory function is not used

 

ba5fnc

5pd

Display refresh rate

 

5if

bp5

Baud rate

Caution The balance may not transmit the data completely at the specified refresh rate, depending on the baud rate or data added to the weighing data such as time, date and ID number.

38

Image 40
Contents GH-120 GH-200 GH-300 GH-202 GH-252 1WMPD4000891CPage Contents Underhook Density Measurement Interface And CommunicationIntroduction About This ManualFeatures ComplianceCompliance With FCC Rules Compliance With EMC DirectivesPage Unpacking the Balance Installing the Balance PrecautionsBefore Use During Use ShockPower Supply After UseDisplay Symbols and Key Operation Key When pressed and released When pressed and heldDisplay Key operationWeighing Units Units101 00001 Counting mode Pcs Percent mode Densitymode Press the Print Changing the UnitsExamples Unit Display Gram Balance displaysUnit setting example Precautions for the weighing operation Basic Operation Gram ModeRead . Display symbols and Key operation before operation WeighingDual Range Weighing range Available minimum display01mg 1mgStoring a sample unit mass Weighing input mode Counting Mode PCSCounting operation Counting mode using the Acai function Mark turns off after Unit mass is improved RepeatPercent Mode % Selecting the percent modeStoring the 100% reference mass Reading the percentageAutomatic Response Adjustment Response AdjustmentAdvise Manual Response Adjustment Key untilDisplayed, then release the key Press Key again quicklyCalibration Calibration GroupAutomatic self calibration due to changes in temperature Automatic Self CalibrationOne-Touch Calibration MarkCalibration Test Using the Internal Mass Press and hold the CAL Key until is CCCalibration Using an External Weight Calibration Test Using an External Weight Key................. The key To cancel the operationKey ..............The key to store the new mass value Even if the AC adapter is removed,Correcting the Internal Mass Value Function Switch and Initialization Permit or InhibitInitializing the Balance Press the Print key To cancel this operation, pressKey to display Setting the Function Table Display symbol and keys Function TableSetting procedure Function table is displayed PressKey to select a class Press Key to enter the class Key to select a item PressSetting example Details of the Function Table Factory settings IntData output Time/Date output ID number output5if Timeout AK, Error codeErCd Interface Ct5Description of the Class Environment, Display Description of the Item Data Output Mode Key modeStream mode Auto print modes a and BInterval memory mode Description of the Item Data FormatStandard format DP Dump print formatKF format MT formatNU numerical format Data Unit Terminator Stable value Unstable valueCSV format Data numberDate TimePositive error Negative error SpaceData Format Examples Stable Unstable OverloadUnits Setting the time Clock and Calendar FunctionOperation Confirming the timeQuitting the operation Setting the dateSet the date Store Setting the ID Number ID Number and GLP ReportGLP Report Calibration report using the internal massKey operation Balance returns to weighing mode after this test Calibration test report using an internal massCommad Calibration report using an external weight Calibration test report using an external weight When using preset target value, proceed to stepTitle block Title block and end blockWeighing data Returns to the weighing mode Data MemoryReleasing Clr or err Press and hold the Print key until the keyData Memory for Weighing Data Storing the weighing dataMemory full Interval memorySetting the function table Parameter settings for each output mode are as followsAdditional parameter settings, as follows Enabling the data memory functionTransmitting all memory data at one time Recalling the memory dataDeleting all memory data at one time Data Memory for Calibration and Calibration Test Storing the calibration and calibration test dataTransmitting the memory data Deleting data stored in memoryUnderhook Density Measurement Formula to obtain the densityPrior to measurement Changing the function table Density of a liquidRelation between the water temperature and density Entering the density of a liquidMeasuring the density Output or store Density display Turns off. lightsPress the modes Key to proceed to otherRS-232C Interface Standard Input and Output InterfaceSub 25 pin assignments Connection to Peripheral Equipment Connection to an AD-8121B printerUsing Windows Communication Tools Software WinCT Using the WinCT software, the balance can do the followingConnection to a computer and the use of WinCT RsComCommands Command ListControl Using CTS and RTS Acknowledge Code and Error CodesSettings Related to RS-232C On command turning on the balance Command ExamplesCommand Re-zeroing the display CAL command Calibration with internal mass Error codeCommand ?PT command Net value Tare value Weighing with a tarePT command to set a target weight of 10 g Zero display Command ExecutionZero Error Codes MaintenanceTreatment of the Balance Sample mass error Unit mass errorAutomatic response adjustment zero error Automatic response adjustment unstable errorOther Display Checking that the balance performs properly Checking the Balance Performance and EnvironmentAsking for Repair Specifications External Dimensions Unit mmOptions and Peripheral Equipment GH-02 USB interfaceGH-08 Ethernet Interface AD-8121B PrinterAD-1653 Density Determination Kit AD-1683 DC Static EliminatorAX-USB-9P-EX USB Converter 9970 g/cm3AX-SW128 Foot switch AX-T-314A-S External key PlugOther options Terms/Index TermsIndex Options 5pd Out5tart Tare PU5eMemo Memo