8.6.Calibration Test Using an External Weight

This function tests the weighing accuracy using an external weight. Calibration test report can be output or stored with "GLP output (info)" (Calibration test does not perform calibration).

1 Connect the AC adapter and warm up the balance for at least one hour with nothing on the weighing pan.

2Press and hold the CAL key until CC out is displayed, then release the key.

3 The balance displays CC 0 .

If the target mass is changed, press the RANGE key and proceed to step 4. A list of usable weights is on page 23.

If current target mass value is used, proceed to step 5.

4 Specify the target mass value as follows:

RANGE key............The key to switch blinking figures.

RE-ZERO(+)key.......The keys to select the target mass or

 

 

(-)key

adjust the mass value. Refer to page 23.

 

MODE

 

 

 

 

key..............The key to store the new mass value.

 

PRINT

 

 

 

 

 

 

Even if the AC adapter is removed, the

 

 

 

 

 

data is maintained in non-volatile memory.

 

 

 

key.................The key

to cancel the operation and

 

CAL

 

 

 

 

 

return to

 

.

 

 

 

 

 

CC 0

Press

and hold

Release

Select

Specify

5 Confirm that there is nothing on the pan and press the PRINT key. The balance measures the zero point and displays the measured value. Do not allow vibration or drafts to affect the balance. The balance displays the target mass value.

Example:

New weight 100.0012 g

6 Place the displayed target mass on the pan and press the

 

PRINT key. The balance measures the target mass and

 

displays the measured value. Do not allow vibration or drafts

 

to affect the balance.

Place weight

7 The balance displays end . Remove the weight from the pan.

8The balance displays glp and outputs or stores "calibration test report. Refer to "11.2. GLP Report" of the function table for details.

9The balance will automatically return to the weighing mode.

Remove

GLP output

27

Image 29
Contents 1WMPD4000891C GH-120 GH-200 GH-300 GH-202 GH-252Page Contents Interface And Communication Underhook Density MeasurementAbout This Manual IntroductionCompliance FeaturesCompliance With FCC Rules Compliance With EMC DirectivesPage Unpacking the Balance Before Use PrecautionsInstalling the Balance Shock During UseAfter Use Power SupplyKey When pressed and released When pressed and held Display Symbols and Key OperationDisplay Key operationUnits Weighing Units101 00001 Changing the Units Counting mode Pcs Percent mode Densitymode Press the PrintExamples Unit Display Gram Balance displaysUnit setting example Basic Operation Gram Mode Precautions for the weighing operationRead . Display symbols and Key operation before operation WeighingWeighing range Available minimum display Dual Range01mg 1mgCounting operation Counting Mode PCSStoring a sample unit mass Weighing input mode Mark turns off after Unit mass is improved Repeat Counting mode using the Acai functionSelecting the percent mode Percent Mode %Storing the 100% reference mass Reading the percentageAdvise Response AdjustmentAutomatic Response Adjustment Key until Manual Response AdjustmentDisplayed, then release the key Press Key again quicklyCalibration Group CalibrationAutomatic Self Calibration Automatic self calibration due to changes in temperatureOne-Touch Calibration MarkPress and hold the CAL Key until is CC Calibration Test Using the Internal MassCalibration Using an External Weight Key................. The key To cancel the operation Calibration Test Using an External WeightKey ..............The key to store the new mass value Even if the AC adapter is removed,Correcting the Internal Mass Value Permit or Inhibit Function Switch and InitializationKey to display Press the Print key To cancel this operation, pressInitializing the Balance Function Table Setting the Function Table Display symbol and keysFunction table is displayed Press Setting procedureKey to select a class Press Key to enter the class Key to select a item PressSetting example Details of the Function Table Int Factory settingsData output Time/Date output ID number outputAK, Error code 5if TimeoutErCd Interface Ct5Description of the Class Environment, Display Key mode Description of the Item Data Output ModeStream mode Auto print modes a and BDescription of the Item Data Format Interval memory modeStandard format DP Dump print formatMT format KF formatNU numerical format Data Unit Terminator Stable value Unstable valueData number CSV formatDate TimeNegative error Space Positive errorData Format Examples Stable Unstable OverloadUnits Clock and Calendar Function Setting the timeOperation Confirming the timeSet the date Store Setting the dateQuitting the operation ID Number and GLP Report Setting the ID NumberKey operation Calibration report using the internal massGLP Report Commad Calibration test report using an internal massBalance returns to weighing mode after this test Calibration report using an external weight When using preset target value, proceed to step Calibration test report using an external weightWeighing data Title block and end blockTitle block Data Memory Returns to the weighing modeReleasing Clr or err Press and hold the Print key until the keyStoring the weighing data Data Memory for Weighing DataMemory full Interval memoryParameter settings for each output mode are as follows Setting the function tableAdditional parameter settings, as follows Enabling the data memory functionDeleting all memory data at one time Recalling the memory dataTransmitting all memory data at one time Storing the calibration and calibration test data Data Memory for Calibration and Calibration TestTransmitting the memory data Deleting data stored in memoryUnderhook Formula to obtain the density Density MeasurementPrior to measurement Changing the function table Density of a liquidEntering the density of a liquid Relation between the water temperature and densityOutput or store Density display Turns off. lights Measuring the densityPress the modes Key to proceed to otherSub 25 pin assignments Standard Input and Output InterfaceRS-232C Interface Connection to an AD-8121B printer Connection to Peripheral EquipmentUsing the WinCT software, the balance can do the following Using Windows Communication Tools Software WinCTConnection to a computer and the use of WinCT RsComCommand List CommandsSettings Related to RS-232C Acknowledge Code and Error CodesControl Using CTS and RTS Command Re-zeroing the display Command ExamplesOn command turning on the balance Error code CAL command Calibration with internal massWeighing with a tare Command ?PT command Net value Tare valueZero Command ExecutionPT command to set a target weight of 10 g Zero display Treatment of the Balance MaintenanceError Codes Unit mass error Sample mass errorAutomatic response adjustment zero error Automatic response adjustment unstable errorOther Display Asking for Repair Checking the Balance Performance and EnvironmentChecking that the balance performs properly Specifications Unit mm External DimensionsGH-02 USB interface Options and Peripheral EquipmentGH-08 Ethernet Interface AD-8121B PrinterAD-1683 DC Static Eliminator AD-1653 Density Determination KitAX-USB-9P-EX USB Converter 9970 g/cm3AX-T-314A-S External key Plug AX-SW128 Foot switchOther options Terms Terms/IndexIndex 5pd Out Options5tart Tare PU5eMemo Memo