Emerson 396RVP instruction manual Section Installation, Installation Through Ball Valve

Page 10

MODEL 396R pH/ORP

SECTION 2.0

 

INSTALLATION

SECTION 2.0

INSTALLATION

2.1UNPACKING AND INSPECTION. Inspect the outside of the carton for any damage. If damage is detected, contact the carrier immediately. Inspect the instrument and hardware. Make sure all items in the packing list are present and in good condition. Notify the factory if any part is missing.

NOTE

If the sensor is to be stored, the protective boot should be filled with either KCl elec- trolyte solution or pH 4.0 buffer solution and replaced on sensor tip until ready to use.

NOTE

Save the original packing cartons and materials as most carriers require proof of damage due to mishandling, etc. Also, if it is necessary to return the instrument to the factory, you must pack the instrument in the same manner as it was received. Refer to Section 8.0 for instructions.

WARNING

Glass electrode must be wetted at all times (in storage and in line) to maximize sensor life.

2.2MECHANICAL INSTALLATION. The Model 396R Sensor may be installed through a weldalet or in a pipe tee or “Y”, as shown in Figure 2-1, when used with a ball valve. Insert the end of the sensor to a depth sufficient to ensure that the glass bulb is continuously wetted by the process fluid. The Model 396R can also be inserted directly into the process without the use of a ball valve for applications not requiring continuous operation during sensor maintenance.

CAUTION

Allow sufficient room for safe retraction and insertion of the sensor. Personnel should have room for stable foot- ing while performing removal or insertion of the sensor.

The sensor must be mounted within 10-90 degrees of the horizontal with the tip pointed downward, thus keeping air bubbles off of the pH sensitive glass bulb. Bubbles settled on the glass bulb disrupt the electrical continuity between the pH sensitive glass and the sil- ver/silver chloride measuring element.

If the retraction version is to be installed without a ball valve follow the installation procedure for insertion service (Section 2.2.2). Perform the following steps for sensor installation through a ball valve:

2.2.1 INSTALLATION THROUGH BALL VALVE.

1.Carefully remove the liquid filled rubber boot which protects the glass electrode and keeps the liquid junction wet during shipping and storage. Discard the liquid and boot. Make sure the lubricated O-ring is in place in the groove inside the male connector on the sensor body.

CAUTION

Buffer solution, in the protective boot, may cause skin or eye irritation.

2.With the male connector on the sensor’s body, insert the sensor into the ball valve until it gently touches the closed valve. The molded electrode guard will protect the glass bulb from breakage.

3.Thread the male connector body tightly into the ball valve assembly. DO NOT tighten the hex nut on the male connector body; doing so would not allow the sensor to be inserted through the ball valve.

4.Pull back hard on the sensor assembly, as if trying to remove the sensor, to be certain that the sensor cannot come free of the ball valve assembly. The built-in retraction stop will butt against the shoul- der of the male connector if properly installed.

CAUTION

The sensor must be captured by the valve assembly and the male connector so that it cannot be blown free by process pressure if mishandled during inser- tion or retraction.

5.After confirming that the sensor assembly is prop- erly secured by the valve assembly, the valve may be opened and the sensor positioned into the process at the desired depth and orientation.

6

Image 10
Contents Retractable pH/ORP Sensors Essential Instructions Special Conditions for safe useModel 396R and 396RVP Table of ContentsList of Tables List of Figures Figure No TitleSection Description and Specifications Features and ApplicationsModel 396RVP Performance and Physical Specifications Model 396RMinimum Conductivity 100 µS/cm Weight/Shipping Weight Measured Range Ordering Information ModelChoose one Commonly Used Accessories for Model 396ROther Accessories for Model 396R BNC Adapter Choose oneCommonly Used Accessories for Model 396RVP Other Accessories for Model 396RVPSection Installation Installation Through Ball ValveSection Typical Mounting Configurations for Model 396R Dimensional Drawing Model 396R with Optional Ball Valve PN 40396R05 D 920 Front Side 396R-21 Back Side 396R Front Side 396R-25 Section Section Section Wiring Model 396R Wiring Model 396R40396R07 40396R012 40396R09 40396R13 40396R06 40396R21 40396R22 14. Wiring Model 396R- -54 to Model Section Wiring Model 396RVP Wiring Model 396RVPModel 81 Wiring Model 2081 Wiring 10. Model 1054 Wiring 14. Model 54/54e Wiring 18. Model Xmt-P-XX-10 Wiring 21. Model 1055-10-22-32 Wiring 396R pH Calibration Recommended two point buffer calibration procedureSection Start UP and Calibration Recommended pH Sensor StandardizationORP of Saturated Quinhydrone Solution Model 396R-21 21 tube Section MaintenanceRetractable Version Model 396R-25 36 tubeCompensation Elements ElementsSection Example of Sensor Tube Replacement Diagnostic Message Section Diagnostics and TroubleshootingTroubleshooting with Diagnostics Description of Problem RemedySensors. For code 01, replace sensor Section Return of Material Warranty RepairNON-WARRANTY Repair GeneralWarranty Return of MaterialSpecifications subject to change without notice