Emerson 396RVP instruction manual Compensation Elements

Page 37

MODEL 396R pH/ORP

SECTION 6.0

 

MAINTENANCE

4.Following the caution above, wash the glass bulb in dilute 5% hydrochloric acid solution and then rinse it thoroughly in tap water. Replace the sen- sor if it cannot be cleaned. If the glass bulb appears clean, proceed to step 5.

5.Buffer calibrate the sensor (Refer to Section 5.0). If the sensor appears to respond sluggishly to pH change, soaking it overnight in a weak acid solution (5% hydrochloric acid) may improve its response. Be sure to follow the CAUTION above and to rinse the sensor’s tip thoroughly with tap water. If the sensor will not calibrate, it must be replaced.

6.4Cleaning Platinum Electrode. The electrode is never exposed to these undesirable compounds. In the event poisoning is suspected, the electrode can be restored to normal operation by simply cleaning the plat- inum electrode with baking soda. Polish it by rubbing it with a damp paper towel and baking soda until a bright, shiny appearance is attained.

6.5Automatic Temperature Compensator. The tem- perature compensator element is temperature sensi- tive and can be checked with an ohmmeter. Resistance increases with temperature.

The 3K element will read 3000 ohms ± 1% at 25°C (77°F) and a Pt-100 will read 110 ohms. Resistance varies with temperature for a 3K and Pt-100 element and can be determined according to Table 6-2 or the following formula:

RT=Ro [l+R1 (T-20)]

Where RT = Resistance

T = Temperature in °C

Refer to Table 6-1 for Ro and R1 values:

TABLE 6-1

Ro and R1 VALUES FOR TEMPERATURE

COMPENSATION ELEMENTS

Temperature

Ro

R1

Compensation Element

 

 

3K

2934

.0045

PT-100

107.7

.00385

 

 

 

TABLE 6-2

TEMPERATURE vs RESISTANCE OF AUTO

T.C. ELEMENTS

 

Resistance

Temperature °C

(Ohms) ±1%

 

3K

PT-100

0

2670

100.0

10

2802

103.8

20

2934

107.7

25

3000

109.6

30

3066

111.5

40

3198

115.4

50

3330

119.2

60

3462

123.1

70

3594

126.9

80

3726

130.8

90

3858

134.6

100

3990

138.5

 

 

 

6.6Sensor Tube Replacement When Used With A Sensor Head Junction Box. Replacement of the retraction versions sensor tube assembly involves the removal and installation of two sets of male connectors: One at the process end of the sensor, and the other at the junction box end (See Figures 6-1,6-2). Refer to Section 6.2 for proper removal of the sensor from process.

1.Remove sensor from process before proceeding. The junction box with attached male connector must be recovered from the old sensor for reuse. Unscrew the junction box cover and set aside. Disconnect electrical connections from printed cir- cuit board inside junction box. Disconnect BNC connector to preamp. Unscrew hex nut (D) from male connector body (C). Separate junction box from used sensor. Set aside.

2.Pry off split ferrule from sensor and set aside for reuse. Remove hex nut (D) and set aside for reuse. Check that the internal O-ring is in place in the male connector body (C) attached to the junc- tion box.

3.Remove hex nut (B) from male connector body (A) at process end of sensor and set aside. Slide the Teflon ferrule and the male connector off sensor in the direction of junction box and set

NOTE

If stainless steel ferrule was used, male connector body (A) will have to be dis- carded with the sensor tube.

33

Image 37
Contents Retractable pH/ORP Sensors Special Conditions for safe use Essential InstructionsModel 396R and 396RVP Table of ContentsList of Tables Figure No Title List of FiguresFeatures and Applications Section Description and SpecificationsMinimum Conductivity 100 µS/cm Weight/Shipping Weight Performance and Physical Specifications Model 396RModel 396RVP Measured RangeModel Ordering InformationOther Accessories for Model 396R Commonly Used Accessories for Model 396RChoose one BNC Adapter Choose oneOther Accessories for Model 396RVP Commonly Used Accessories for Model 396RVPInstallation Through Ball Valve Section InstallationSection Typical Mounting Configurations for Model 396R Dimensional Drawing Model 396R with Optional Ball Valve PN 40396R05 D 920 Front Side 396R-21 Back Side 396R Front Side 396R-25 Section Section Wiring Model 396R Section Wiring Model 396R40396R07 40396R012 40396R09 40396R13 40396R06 40396R21 40396R22 14. Wiring Model 396R- -54 to Model Wiring Model 396RVP Section Wiring Model 396RVPModel 81 Wiring Model 2081 Wiring 10. Model 1054 Wiring 14. Model 54/54e Wiring 18. Model Xmt-P-XX-10 Wiring 21. Model 1055-10-22-32 Wiring Section Start UP and Calibration Recommended two point buffer calibration procedure 396R pH Calibration Recommended pH Sensor StandardizationORP of Saturated Quinhydrone Solution Retractable Version Section MaintenanceModel 396R-21 21 tube Model 396R-25 36 tubeElements Compensation ElementsSection Example of Sensor Tube Replacement Troubleshooting with Diagnostics Section Diagnostics and TroubleshootingDiagnostic Message Description of Problem RemedySensors. For code 01, replace sensor NON-WARRANTY Repair Warranty RepairSection Return of Material GeneralReturn of Material WarrantySpecifications subject to change without notice