Emerson 396RVP instruction manual Section Start UP and Calibration, 396R pH Calibration

Page 34

MODEL 396R pH/ORP

SECTION 5.0

 

START UP AND CALIBRATION

SECTION 5.0

START UP AND CALIBRATION

5.1START UP. To obtain best accuracy, the sensor must be calibrated as a loop with the analyzer. Please refer to the respective analyzer/transmitter instruction manual for proper calibration procedures.

5.2396R pH CALIBRATION

1.After a temporary connection is established between the sensor and the instrument, a buffer calibration may be performed.

2.Consult appropriate pH/ORP analyzer or transmitter instruction manual for specific calibration and standardization procedures or see below for recommended two point buffer calibration procedure.

5.2.1Recommended two point buffer calibration procedure:

Select two stable buffer solutions, preferably pH 4.0 and 10.0 (pH buffers other than pH 4.0 and pH 10.0 can be used as long as the pH values are at least two pH units apart).

Note: A pH 7 buffer solution reads a mV value of approx. zero, and pH buffers read approx. +/- 59.1 mV for each pH unit above or below pH 7. Check the pH buffer manufacturer spec- ifications for millivolt values at various temper- atures since it may affect the actual value of the buffer solution mV/pH value.

1.Immerse sensor in the first buffer solution. Allow sensor to adjust to the buffer temperature (to avoid errors due to temperature differences between the buffer solution and sensor temperature) and wait for reading to stabilize. Value of buffer can now be acknowledged by analyzer/transmitter

2.Once the first buffer has been acknowledged by the analyzer/transmitter, rinse the buffer solution off of the sensor with distilled or deionized water.

3.Repeat steps 1 and 2 using the second buffer solution.

4.Once the analyzer/transmitter has acknowledged both buffer solutions, a sensor slope (mV/pH) is established (the slope value can be found within the analyzer/transmitter).

5.The slope value should read about 59.1 mV/pH for a new sensor and will decrease over time to approximately 47 - 49 mV/pH. Once the slope reads below the 47-49 mV/pH range, a new sensor should be installed to maintain accurate readings.

5.2.2 Recommended pH Sensor Standardization

For maximum accuracy, the sensor can be standard- ized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process. Standardization accounts for the sensor junction potential and other interferences. Standardization will not change the sen- sor’s slope but will simply adjust the analyzers reading to match that of the known process pH.

1.While obtaining a process solution sample (it is recommended that the sample is taken close to the sensor), record the pH value that is shown on the analyzer/transmitter display.

2.Measure and record the pH of the process solution sample with a another temperature compensated, calibrated pH instrument. For best results, standardization should be performed at the process temperature.

3.Adjust the analyzer/transmitter value to the standardized value.

30

Image 34
Contents Retractable pH/ORP Sensors Essential Instructions Special Conditions for safe useModel 396R and 396RVP Table of ContentsList of Tables List of Figures Figure No TitleSection Description and Specifications Features and ApplicationsModel 396RVP Performance and Physical Specifications Model 396RMinimum Conductivity 100 µS/cm Weight/Shipping Weight Measured RangeOrdering Information ModelChoose one Commonly Used Accessories for Model 396ROther Accessories for Model 396R BNC Adapter Choose oneCommonly Used Accessories for Model 396RVP Other Accessories for Model 396RVPSection Installation Installation Through Ball ValveSection Typical Mounting Configurations for Model 396R Dimensional Drawing Model 396R with Optional Ball Valve PN 40396R05 D 920 Front Side 396R-21 Back Side 396R Front Side 396R-25 Section Section Section Wiring Model 396R Wiring Model 396R40396R07 40396R012 40396R09 40396R13 40396R06 40396R21 40396R22 14. Wiring Model 396R- -54 to Model Section Wiring Model 396RVP Wiring Model 396RVPModel 81 Wiring Model 2081 Wiring 10. Model 1054 Wiring 14. Model 54/54e Wiring 18. Model Xmt-P-XX-10 Wiring 21. Model 1055-10-22-32 Wiring 396R pH Calibration Recommended two point buffer calibration procedureSection Start UP and Calibration Recommended pH Sensor StandardizationORP of Saturated Quinhydrone Solution Model 396R-21 21 tube Section MaintenanceRetractable Version Model 396R-25 36 tubeCompensation Elements ElementsSection Example of Sensor Tube Replacement Diagnostic Message Section Diagnostics and TroubleshootingTroubleshooting with Diagnostics Description of Problem RemedySensors. For code 01, replace sensor Section Return of Material Warranty RepairNON-WARRANTY Repair GeneralWarranty Return of MaterialSpecifications subject to change without notice