Emerson 396RVP instruction manual Example of Sensor Tube Replacement

Page 39

MODEL 396R pH/ORP

SECTION 6.0

 

MAINTENANCE

FIGURE 6-1. Example of Sensor Tube Replacement

FIGURE 6-2. Male Connector Tightening Diagram

35

Image 39
Contents Retractable pH/ORP Sensors Special Conditions for safe use Essential InstructionsTable of Contents Model 396R and 396RVPList of Tables Figure No Title List of FiguresFeatures and Applications Section Description and SpecificationsMeasured Range Performance and Physical Specifications Model 396RMinimum Conductivity 100 µS/cm Weight/Shipping Weight Model 396RVPModel Ordering InformationBNC Adapter Choose one Commonly Used Accessories for Model 396ROther Accessories for Model 396R Choose oneOther Accessories for Model 396RVP Commonly Used Accessories for Model 396RVPInstallation Through Ball Valve Section InstallationSection Typical Mounting Configurations for Model 396R Dimensional Drawing Model 396R with Optional Ball Valve PN 40396R05 D 920 Front Side 396R-21 Back Side 396R Front Side 396R-25 Section Section Wiring Model 396R Section Wiring Model 396R40396R07 40396R012 40396R09 40396R13 40396R06 40396R21 40396R22 14. Wiring Model 396R- -54 to Model Wiring Model 396RVP Section Wiring Model 396RVPModel 81 Wiring Model 2081 Wiring 10. Model 1054 Wiring 14. Model 54/54e Wiring 18. Model Xmt-P-XX-10 Wiring 21. Model 1055-10-22-32 Wiring Recommended pH Sensor Standardization Recommended two point buffer calibration procedureSection Start UP and Calibration 396R pH CalibrationORP of Saturated Quinhydrone Solution Model 396R-25 36 tube Section Maintenance Retractable Version Model 396R-21 21 tubeElements Compensation ElementsSection Example of Sensor Tube Replacement Description of Problem Remedy Section Diagnostics and TroubleshootingTroubleshooting with Diagnostics Diagnostic MessageSensors. For code 01, replace sensor General Warranty RepairNON-WARRANTY Repair Section Return of MaterialReturn of Material WarrantySpecifications subject to change without notice