Emerson 396RVP instruction manual Sensors. For code 01, replace sensor

Page 41

MODEL 396R pH/ORP

SECTION 7.0

 

DIAGNOSTICS AND TROUBLESHOOTING

7.2TROUBLESHOOTING WITHOUT DIAGNOSTICS. Table 7-2 lists common problems, causes and remedies typically encountered in process measurement.

TABLE 7-2. Troubleshooting without Diagnostics

Problem

Probable Cause

 

Remedy

 

 

 

 

 

Meter reads off scale. (Display

Defective preamplifier.

 

Replace preamplifier (for code 02

reads overrange).

 

 

sensors). For code 01, replace sensor.

 

 

 

 

 

T.C. element shorted.

 

Check T.C. element as instructed

 

 

 

in Section 6.5 and

 

 

 

replace sensor if defective.

 

 

 

 

 

Sensor not in process or sample

 

Make sure sensor is in process with

 

stream is low.

 

sufficient sample stream (refer to

 

 

 

Section 2.0 for installation details).

 

 

 

 

 

Open glass electrode.

 

Replace sensor.

 

 

 

 

 

Reference element open - no contact.

 

Replace sensor.

 

 

 

 

Display reads between 3 and 6 pH

Electrode cracked.

 

Replace sensor.

regardless of actual pH of solution

 

 

 

or sample.

 

 

 

 

 

 

 

Meter or display indication swings

T.C. element shorted.

 

Check T.C. element as instructed

or jumps widely in AUTO T.C. Mode.

 

 

in Section 6.5 and replace

 

 

 

sensor if defective.

 

 

 

 

Span between buffers extremely

T.C. element open.

 

Check T.C. element as instructed

short in AUTO T.C. Mode.

 

 

in Section 6.5 and replace sensor

 

 

 

if defective.

 

 

 

 

Sluggish or slow meter indication

Electrode coated.

 

Clean sensor as instructed in

for real changes in pH level.

 

 

Sections 6.3 or 6.4. Replace

 

 

 

sensor if cracked.

 

 

 

 

 

Electrode defective.

 

Replace sensor.

 

 

 

 

Transmitter cannot be standardized.

Electrode coated or cracked.

 

Clean Sensor as instructed in

 

 

 

Sections 6.3 or 6.4 Replace

 

 

 

sensor if cracked.

 

 

 

 

 

Defective preamplifier.

 

Replace preamplifier.

 

 

 

 

Transmitter short spans between

Aged glass electrode or high

 

Replace sensor.

two different buffer values.

temperature exposure.

 

 

 

 

 

 

 

Electrode coated .

 

Clean Sensor as instructed in

 

 

 

Sections 6.3 or 6.4. Replace

 

 

 

sensor if cracked.

 

 

 

 

37

Image 41
Contents Retractable pH/ORP Sensors Special Conditions for safe use Essential InstructionsList of Tables Table of ContentsModel 396R and 396RVP Figure No Title List of FiguresFeatures and Applications Section Description and SpecificationsMinimum Conductivity 100 µS/cm Weight/Shipping Weight Performance and Physical Specifications Model 396RModel 396RVP Measured RangeModel Ordering InformationOther Accessories for Model 396R Commonly Used Accessories for Model 396RChoose one BNC Adapter Choose oneOther Accessories for Model 396RVP Commonly Used Accessories for Model 396RVPInstallation Through Ball Valve Section InstallationSection Typical Mounting Configurations for Model 396R Dimensional Drawing Model 396R with Optional Ball Valve PN 40396R05 D 920 Front Side 396R-21 Back Side 396R Front Side 396R-25 Section Section Wiring Model 396R Section Wiring Model 396R40396R07 40396R012 40396R09 40396R13 40396R06 40396R21 40396R22 14. Wiring Model 396R- -54 to Model Wiring Model 396RVP Section Wiring Model 396RVPModel 81 Wiring Model 2081 Wiring 10. Model 1054 Wiring 14. Model 54/54e Wiring 18. Model Xmt-P-XX-10 Wiring 21. Model 1055-10-22-32 Wiring Section Start UP and Calibration Recommended two point buffer calibration procedure396R pH Calibration Recommended pH Sensor StandardizationORP of Saturated Quinhydrone Solution Retractable Version Section MaintenanceModel 396R-21 21 tube Model 396R-25 36 tubeElements Compensation ElementsSection Example of Sensor Tube Replacement Troubleshooting with Diagnostics Section Diagnostics and TroubleshootingDiagnostic Message Description of Problem RemedySensors. For code 01, replace sensor NON-WARRANTY Repair Warranty RepairSection Return of Material GeneralReturn of Material WarrantySpecifications subject to change without notice