Emerson 396RVP Section Description and Specifications, Features and Applications

Page 5

MODEL 396R pH/ORP

SECTION 1.0

 

DESCRIPTION AND SPECIFICATIONS

SECTION 1.0

DESCRIPTION AND SPECIFICATIONS

1.1 FEATURES AND APPLICATIONS

The Model 396R and 396RVP Sensors are specifically designed for improved life in harsh, dirty applications where a separate sample stream is difficult to provide and greater insertion depths are required. Model 396R is designed for use with a 1-1/4 in. or 1-1/2 in. ball valve for hot tap installa- tion. The Model 396R is constructed of molded polypropylene housed in a titanium tube with EPDM seals to provide maxi- mum chemical resistance2,3.

Model 396R also features a titanium solution ground for advanced sensor diagnostics when used with the Models 1055, Xmt-P, 54e, 81, 3081, or 5081 pH/ORP Analyzer/ Transmitter. Advanced sensor diagnostics provide preventa- tive maintenance by notifying the operator for replacement and cleaning of an aged or fouled sensor for continuous optimum performance.

The sensor also features a shrouded tip for protection from breakage while allowing process to flow by the glass elec- trode for accurate and reliable pH measurement.

The Model 396R is available without an integral preamplifier only and 15 ft or 9.5 in. of integral high quality 9 conductor cable. The preamplifier must be housed in a remote location or in a J-Box kit for attachment at the rear, cable end of the sensor (order separately) or integral to the Analyzer/

Transmitter. The Model 396R is compatible with all Rosemount Analytical and various other manufacturers instruments.

The entire line of TUpH model sensors now incorporate the new SILCORE1 technology contaminant barrier. This triple-seal barrier prevents moisture and material impurities from migrating to the pH sensor’s reference electrode’s metal lead wire. By preventing these contaminants from compro- mising the integrity of the pH measurement, sensor life is increased, especially at higher temperatures where increased migrations occur. In addition, the SILCORE tech- nology provides added protection against sensor failure due to vibrations and shock by transferring damaging energy away from the glass-to-metal seal.

Model 396RVP: Rosemount Analytical has recently released Model 396RVP. This model has identical performance and physical specifications to the Model 396R (see Section 1.2) with the following exception: the Model 396RVP has a Variopol (VP) connector on the back end of the sensor in place of a cable.

1 Silcore is a trademark of Rosemount Analytical.

FIGURE 1-1. Cross Section Diagram of the TUpH Reference Technology

All TUpH sensors are designed with a large area reference junction, helical reference pathway, and an AccuGlass pH

glass bulb. This sensor technology ensures superior performance while only requiring minimal maintenance.

1

Image 5
Contents Retractable pH/ORP Sensors Special Conditions for safe use Essential InstructionsList of Tables Table of ContentsModel 396R and 396RVP Figure No Title List of FiguresFeatures and Applications Section Description and SpecificationsMinimum Conductivity 100 µS/cm Weight/Shipping Weight Performance and Physical Specifications Model 396RModel 396RVP Measured RangeModel Ordering InformationOther Accessories for Model 396R Commonly Used Accessories for Model 396RChoose one BNC Adapter Choose oneOther Accessories for Model 396RVP Commonly Used Accessories for Model 396RVPInstallation Through Ball Valve Section InstallationSection Typical Mounting Configurations for Model 396R Dimensional Drawing Model 396R with Optional Ball Valve PN 40396R05 D 920 Front Side 396R-21 Back Side 396R Front Side 396R-25 Section Section Wiring Model 396R Section Wiring Model 396R40396R07 40396R012 40396R09 40396R13 40396R06 40396R21 40396R22 14. Wiring Model 396R- -54 to Model Wiring Model 396RVP Section Wiring Model 396RVPModel 81 Wiring Model 2081 Wiring 10. Model 1054 Wiring 14. Model 54/54e Wiring 18. Model Xmt-P-XX-10 Wiring 21. Model 1055-10-22-32 Wiring Section Start UP and Calibration Recommended two point buffer calibration procedure396R pH Calibration Recommended pH Sensor StandardizationORP of Saturated Quinhydrone Solution Retractable Version Section MaintenanceModel 396R-21 21 tube Model 396R-25 36 tubeElements Compensation ElementsSection Example of Sensor Tube Replacement Troubleshooting with Diagnostics Section Diagnostics and TroubleshootingDiagnostic Message Description of Problem RemedySensors. For code 01, replace sensor NON-WARRANTY Repair Warranty RepairSection Return of Material GeneralReturn of Material WarrantySpecifications subject to change without notice