MODEL 396R pH/ORP

SECTION 3.0

 

WIRING MODEL 396R

SECTION 3.0

WIRING MODEL 396R

WIRING MODEL 396R.

Make electrical connections as shown on Figures 3-1 through 3-15 using the following guidelines. For wiring Model 396RVP, see Section 4.0.

1.Pay particular attention to the analyzer or trans- mitter model number when following details on the wiring diagrams to ensure that the connections are made to the proper terminals.

2.Use Rosemount custom cable Part Number 9200273 for interconnect.

3.The maximum distance from the sensor to the analyzer is 15 ft without an integral preamplifier.

4.Signal cable should be run in a dedicated conduit and should be kept away from AC power lines.

NOTE

For maximum EMI/RFI protection when wiring from the sensor to the junction box, the outer braid of the sensor should be connected to the outer braided shield of the extension cable. The outer braid of the extension cable to the instrument must be terminated at earth ground or by using an appropriate metal cable gland fitting, that provides a secure connection to the instrument cable.

WARNING: IF INNER BLACK CONDUCTIVE SHEATH IS IN CONTACT WITH THE EXPOSED LEADS, OR IS NOT PREPARED PROPERLY, IT MAY CAUSE AN ELECTRICAL SHORT.

DWG. NO.

40396R24

REV.

A

FIGURE 3-1. Cable Preparation Instructions (PN 9200274)

14

Image 18
Contents Retractable pH/ORP Sensors Essential Instructions Special Conditions for safe useTable of Contents Model 396R and 396RVPList of Tables List of Figures Figure No TitleSection Description and Specifications Features and ApplicationsModel 396RVP Performance and Physical Specifications Model 396RMinimum Conductivity 100 µS/cm Weight/Shipping Weight Measured RangeOrdering Information ModelChoose one Commonly Used Accessories for Model 396ROther Accessories for Model 396R BNC Adapter Choose oneCommonly Used Accessories for Model 396RVP Other Accessories for Model 396RVPSection Installation Installation Through Ball ValveSection Typical Mounting Configurations for Model 396R Dimensional Drawing Model 396R with Optional Ball Valve PN 40396R05 D 920 Front Side 396R-21 Back Side 396R Front Side 396R-25 Section Section Section Wiring Model 396R Wiring Model 396R40396R07 40396R012 40396R09 40396R13 40396R06 40396R21 40396R22 14. Wiring Model 396R- -54 to Model Section Wiring Model 396RVP Wiring Model 396RVPModel 81 Wiring Model 2081 Wiring 10. Model 1054 Wiring 14. Model 54/54e Wiring 18. Model Xmt-P-XX-10 Wiring 21. Model 1055-10-22-32 Wiring 396R pH Calibration Recommended two point buffer calibration procedureSection Start UP and Calibration Recommended pH Sensor StandardizationORP of Saturated Quinhydrone Solution Model 396R-21 21 tube Section MaintenanceRetractable Version Model 396R-25 36 tubeCompensation Elements ElementsSection Example of Sensor Tube Replacement Diagnostic Message Section Diagnostics and TroubleshootingTroubleshooting with Diagnostics Description of Problem RemedySensors. For code 01, replace sensor Section Return of Material Warranty RepairNON-WARRANTY Repair GeneralWarranty Return of MaterialSpecifications subject to change without notice