Chapter 4, Troubleshooting
Errors While Attempting Measurements
Errors While Attempting Measurements
CDMA Measurement not Correlating
NOTE:
Problem:
If your Test System is not set up correctly for some of the tests, you may see the following message posted at the top of the display:
Cannot correlate to input signal. Check setup.
The error message can appear while in many of the menus, but it will affect the following main menu selections:
•Go to Code Domain Analyzer Screen
•Measure Rho Parameters
For an
Possible causes and solutions:
1Base Station Channel entry.
The Software tunes the Test System to the transmitter frequency based on your entry for Base Station Channel. The entry must match the channel your base station is set to transmit on. Verify the channel number entry in the Base Station Channel field on the main menu screen. If you want to verify the frequency the Test System is tuned to, see the TX Freq setting at the bottom of the main menu screen.
2PN Offset entry.
For proper correlation, you must enter the PN Offset value that matches that of the transmitter being tested. Check the entry in the main menu level. If you do not know the PN offset, see "Search for PN Offset" on page 44
3Missing an
You must have access to the