O2 Innovations pmn Example Display Setting Battery Type, Example Display of Set Operation Menu

Page 20

TARGET Series

ENMET Corporation

6.4.6 Set Battery Type

From the Advanced Maintenance menu, press the MENU button five times to put the asterisk next to the Set Battery line. Press the SELECT button to enter the Set Battery screen. Use the MENU button alternate between NiMH (rechargeable) and Alkaline. Press the SELECT button to choose the desired battery type. Setting the proper battery type is important to achieve maximum life out of the battery pack.

WARNING: Substitution of batteries or other components may compromise the intrinsic safety of the instrument.

Set Password

*Set Battery

Set Op. Menu Type

Exit

SELECT

Set BATTERY TYPE Select type: NiMH

Example: Display Setting Battery Type

6.4.7 Set Operation Menu

From the Advanced Maintenance menu, press the MENU button six times to put the asterisk next to the Set Op. menu type line. Press the SELECT button to enter the Set Op. menu type screen. Use the MENU button alternate between Basic and Advanced. Press the SELECT button to choose the desired operational menu. The Basic menu type is the default menu.

Set Password

Set Battery

*Set Op. Menu Type

Exit

SELECT

Set OP MENU TYPE Select type: Basic

Example: Display of Set Operation Menu

6.4.8 Internal Pump Calibration

For instruments with internal sampling pumps, a calibration procedure is required whenever a pump is replaced or when calibration is performed on the instrument sensors. It is also recommended to perform pump calibration if any abnormal pump operation is observed.

NOTE: The calibration or sampling shield should not be attached to the instrument during pump calibration. ￿Press the menu button to move the asterisk next to the Pump Calibration prompt.

￿Press the select button to initiate the procedure.

￿The pump will automatically come on. The instrument samples the pump signal for one minute. At the end of the one minute, the instrument is ready for the blocked flow measurement.

￿Put your finger over one of the two sampling holes on the back of the sensor housing.

￿Press the select button. The instrument samples the blocked flow condition for 10 seconds and then stores the reading. See figure 9, for location of sampling holes.

The low flow alarm is set automatically by the instrument based on the free flow and blocked flow readings. No further adjustments are necessary.

Exit the Maintenance Menu when maintenance operations are complete.

17

Image 20
Contents Target Series Enmet CorporationTable of Contents List of Figures and Tables Unpack Regarding Damaged ShipmentsTarget Features Visual Alarms Display Charging Port Horn ON/OFF Button Select Button Menu Button GasketExample Power Down Display Instrument OperationLocations and Data logging Turning the Instrument On and OffExample Display, Charging Complete Charging the BatteryExample Display, Battery Charging Battery Life AlarmsInstantaneous Gas Alarm STEL/TWA AlarmsExample Display, Auto-Zero from Pop-Up Menu Pop-Up MenuExample Pop-Up Menu Example Display, Combustible Gas Example Display of Calibration StatusProgrammed Flammable Gas Coefficients Example Display, Internal Pump Status Internal Sampling PumpDate/Time/Battery Information Example Display of Date, Time, Battery LevelAlarms Values Example Display of Alarm ValuesStel Values Min/Max ValuesTWA Values Maintenance Operations Standard Maintenance MenuRecommendations CalibrationZero Adjust Applying Span GasCalibration Adapter, BRH 04834-002 Calibration Adapter, Reactive GasesAdvanced Maintenance Menu, Software 4.94 version Example Display of Advanced Maintenance MenuExample Display of Advanced Calibration Menu Example Display of BRH CalibrationExample Display Changing Password Example Display of Set Alarm MenuExample Display while Setting Clock and Chirp interval Example Display Setting Battery Type Set Battery TypeExample Display of Set Operation Menu Set Operation Menu4a.1 Calibration 4a Advanced Maintenance Menu, Software 6.0 version4a.2 Setting Instantaneous Alarms 4a.3 Enable Sensors 4a.4 Internal Pump CalibrationReplacing Battery Pack PCBReplacing Sensors Remove shorting clipInstrument Information Enmet Part No DescriptionTarget Specifications Element SpecificationDefault Configuration Default Alarm and ConfigurationsDefault Alarm Basic Operation Menu Flow DiagramsTarget Menu Quick Reference Enter any calibration procedure from above Advanced Operation Menu BRH Advanced Maintenance Menu, Software 6.0 version Troubleshooting Symptom Possible Cause ActionWarranty Returning an Instrument for Repair PO Box Fairfield Court Ann Arbor, Michigan 734.761.1270 FaxRepair Return Form