O2 Innovations pmn manual Example Display of Set Alarm Menu, Example Display Changing Password

Page 19

TARGET Series

ENMET Corporation

button at the Exit line to go back to the Maintenance menu. STEL and TWA alarms are preset at the factory and can only be changed using the TARGETTrakker software package.

Calibrate

*Set Alarms

Enable Sensors

Set Clock/Chirp

SELECT

SET ALARMS

*O2 19.5 / 23.5 % CO 35 ppm

H2S 10 ppm

Example: Display of Set Alarm Menu

6.4.3 Enable Sensors

From the Advanced Maintenance menu, press the MENU button twice once to put the asterisk next to the Enable Sensors line. Press the SELECT button to enter the Enable Sensors screen. Use the MENU button to move the asterisk next to the gas whose alarm needs to be changed. Press the SELECT button to toggle the sensor ON or OFF. Press SELECT at the Exit line to go back to the Maintenance menu.

Calibrate

Set Alarms

*Enable Sensors

Set Clock/Chirp

SELECT

ENABLE SENSORS *O2 ON

CO ON

H2S ON

6.4.4 Set Clock/Chirp

From the Advanced Maintenance menu, press the MENU button three times to put the asterisk next to the Set Clock/Chirp line. Press the SELECT button to enter the Set Clock/Chirp screen. Use the MENU button to move the asterisk next to the value that needs to be changed. Press the SELECT button to move a cursor to the first digit. The MENU button is used to increment the digit and the SELECT button to move to the next digit.

Time must be entered in a 24 hour format (i.e. 4:30 pm is entered as 16:30). The confidence chirp sounds at interval entered in this screen. Factory default value is 120 seconds.

Press the SELECT button at the Exit line to go back to the Maintenance menu.

Calibrate

Set Alarms

Enable Sensors

*Set Clock/Chirp

SELECT

*M/D/Y: 05/22/99 Time (24Hr): 18:53 Chirp (sec): 120 Exit

Example: Display while Setting Clock and Chirp interval

6.4.5 Set Password

From the Advanced Maintenance menu, press the MENU button four times to put the asterisk next to the Set Password line. Press the SELECT button to enter the Set Password screen. Use the MENU button to increment the digit and the SELECT button to move to the next digit. Any four digit password from 0000 to 9999 may be entered by this operation.

*Set Password

Set Battery

Set Op. Menu Type

Exit

SELECT

CHANGE PASSWORD Enter new value: 1270

Example: Display Changing Password

16

Image 19
Contents Enmet Corporation Target SeriesTable of Contents List of Figures and Tables Regarding Damaged Shipments UnpackTarget Features Horn ON/OFF Button Select Button Menu Button Gasket Visual Alarms Display Charging PortTurning the Instrument On and Off Example Power Down DisplayInstrument Operation Locations and Data loggingExample Display, Battery Charging Charging the BatteryExample Display, Charging Complete STEL/TWA Alarms Battery LifeAlarms Instantaneous Gas AlarmExample Pop-Up Menu Pop-Up MenuExample Display, Auto-Zero from Pop-Up Menu Programmed Flammable Gas Coefficients Example Display of Calibration StatusExample Display, Combustible Gas Internal Sampling Pump Example Display, Internal Pump StatusExample Display of Alarm Values Date/Time/Battery InformationExample Display of Date, Time, Battery Level Alarms ValuesTWA Values Min/Max ValuesStel Values Calibration Maintenance OperationsStandard Maintenance Menu RecommendationsApplying Span Gas Zero AdjustCalibration Adapter, Reactive Gases Calibration Adapter, BRH 04834-002Example Display of BRH Calibration Advanced Maintenance Menu, Software 4.94 versionExample Display of Advanced Maintenance Menu Example Display of Advanced Calibration MenuExample Display while Setting Clock and Chirp interval Example Display of Set Alarm MenuExample Display Changing Password Set Operation Menu Example Display Setting Battery TypeSet Battery Type Example Display of Set Operation Menu4a.2 Setting Instantaneous Alarms 4a Advanced Maintenance Menu, Software 6.0 version4a.1 Calibration 4a.4 Internal Pump Calibration 4a.3 Enable SensorsPCB Replacing Battery PackRemove shorting clip Replacing SensorsEnmet Part No Description Instrument InformationElement Specification Target SpecificationsDefault Alarm Default Alarm and ConfigurationsDefault Configuration Target Menu Quick Reference Flow DiagramsBasic Operation Menu Enter any calibration procedure from above Advanced Operation Menu BRH Advanced Maintenance Menu, Software 6.0 version Symptom Possible Cause Action TroubleshootingWarranty PO Box Fairfield Court Ann Arbor, Michigan 734.761.1270 Fax Returning an Instrument for RepairRepair Return Form