TARGET Series | ENMET Corporation |
Sensor end cap w/cal shield
Regulator
Humidifier Bowl
Cylinder,
Calibration Gas
Example of Storage
Calibration Adapter, BRH
Sensor end cap w/cal shield
Sensor end cap
w/cal shield
Humidifier Bowl | Regulator | Regulator |
|
|
Top View of Humidifier Bowl
←To Sensor | To Regulator |
|
Cylinder, Calibration Gas
Calibration Adapter, BRH
04834-002
Cylinder, Calibration Gas
Calibration Adapter, Reactive Gases
Figure 8: Various Calibration Adapter Attachments
Apply the span gas and press the SELECT button to initiate the sample count down timer. Count down is shown on the lowest line of the display, at the end of the count down the span reading is stored in memory. Remove the span gas and the calibration shield. Calibration of this sensor is complete.
If the sensor is within acceptable sensitivity parameters, OK is displayed next to the gas momentarily before the display returns to the Calibration menu.
If there is a problem, Cal error is displayed next to the appropriate channel. See section 10.0 for troubleshooting.
CAL: O2/CO/H2S/Mth Sampling
Sample in 120 sec.
Example:
Display during and after Calibration
O2: | OK |
CO: | Cal error |
H2S: | OK |
Mth: | OK |
|
|
Repeat the above procedure for each individual sensor as required. If a calibration error occurs it will also be displayed as a general condition on the Status Line of the Main Gas Display and can be viewed in the Cal Status screen of the
Exit the Calibration menu when completed. Remove span gas and calibration shield. Then exit the Standard Maintenance menu.
14