6.3.2 Zero Adjust

The calibration zero procedure is activated by pressing the SELECT button again. This starts a 30 second count down timer, at the end of which, the zero reading is stored in memory. The next screen displays the span gas concentration.

CAL: O2/CO/H2S/Mth Sampling CLEAN AIR MENU:Exit

Sample in 30 sec.

Example:

Display of Calibration Count Down

CAL: O2/CO/H2S/Mth Apply Multi-Blend SEL: frwd MENU:Exit

NOTE: The oxygen sensor requires only the initial 30 second clean air procedure.

6.3.3 Applying Span Gas

Attach the calibration adapter to the cylinder of span gas and calibration shield. See figure 8.

CAUTION: Some gas require that the Target Sampling Shield and flow demand regulator be used for calibration. See Section 8.0, Table 3 and 4.

I

BRH L

:

F A SENSOR IS INSTALLED IN

OCATION

￿Place instrument on charge over night.

 

￿The BRH sensor should be calibrated before any others. If the BRH sensor is installed in the TARGET, it will

become contaminated if exposed to the combination or test gas and must be calibrated separately.

￿If the BRH sensor is being calibrated with the 04834-002 calibration fixture, fill the humidifier bowl approximately

1/3 full with clean, tap water. The tube should NOT go under the water line. No bubbles should be visible when

 

gas is flowing. See figure 8.

NOTE: The arrow on the humidifier bowl points in the opposite direction of airflow.

￿If the BRH sensor is being calibrated with the 04834-005 calibration fixture, fill the humidifier bowl to level

 

indicated. Use the storage case to support the humidifier while in use. See figure 8.

NOTE: The arrow on the humidifier bowl points in the same direction of airflow. 04834-005 calibration fixture

 

 

supplied starting November 2002.

￿Failure to follow both of the BRH steps will cause an inaccurate calibration.

F

A

:

OR

 

LL OTHER SENSORS INSTALLED

￿ Use the reactive gases calibration adapter which does not have a humidifier bowl. For CO, O2, Methane sensor

calibration these sensors may be calibrated simultaneously or individually using the combination calibration gas and reactive gases calibration adapter. See figure 8.

NOTE: When calibrating CO, O2 and CH4, with individual gas cylinders, the BRH calibration adapter should be used.

￿Failure to do so will cause an inaccurate calibration.

SPAN gas concentrations are not adjustable in this calibration procedure. SPAN gas concentrations are, by design, the concentrations available with ENMET calibration kits. See Section 8.0 for the default concentrations. If other concentrations are to be used for calibration, the calibration procedure in the Advanced Maintenance menu must be followed.

NOTE: If the span gas value has been previously changed in the Advanced Maintenance menu, that change is carried over to this procedure.

Attach the calibration shield to the sensor end cap of the instrument. See Figure 8A. Guide the flap of the shield into the groove on the bottom of the end cap and bring the top of the shield towards the top of the end cap. Tighten the thumb screw into the threaded hole in the sensor end cap.

TOP

Threaded hole for thumb screw

Sensor

End Cap

BOTTOM

Groove

Sensor

End Cap

 

Thumb screw

 

Calibration Shield

 

Sensor End Cap

Thumb screw

with Calibration

Calibration Shield

Shield attached

 

Flap of Shield

 

Figure 8a: Calibration Shield Attachment

13

Image 16
Contents Target Series Enmet CorporationTable of Contents List of Figures and Tables Unpack Regarding Damaged ShipmentsTarget Features Visual Alarms Display Charging Port Horn ON/OFF Button Select Button Menu Button GasketExample Power Down Display Instrument OperationLocations and Data logging Turning the Instrument On and OffExample Display, Battery Charging Charging the BatteryExample Display, Charging Complete Battery Life AlarmsInstantaneous Gas Alarm STEL/TWA AlarmsExample Pop-Up Menu Pop-Up MenuExample Display, Auto-Zero from Pop-Up Menu Programmed Flammable Gas Coefficients Example Display of Calibration StatusExample Display, Combustible Gas Example Display, Internal Pump Status Internal Sampling PumpDate/Time/Battery Information Example Display of Date, Time, Battery LevelAlarms Values Example Display of Alarm ValuesTWA Values Min/Max ValuesStel Values Maintenance Operations Standard Maintenance MenuRecommendations CalibrationZero Adjust Applying Span GasCalibration Adapter, BRH 04834-002 Calibration Adapter, Reactive GasesAdvanced Maintenance Menu, Software 4.94 version Example Display of Advanced Maintenance MenuExample Display of Advanced Calibration Menu Example Display of BRH CalibrationExample Display while Setting Clock and Chirp interval Example Display of Set Alarm MenuExample Display Changing Password Example Display Setting Battery Type Set Battery TypeExample Display of Set Operation Menu Set Operation Menu4a.2 Setting Instantaneous Alarms 4a Advanced Maintenance Menu, Software 6.0 version4a.1 Calibration 4a.3 Enable Sensors 4a.4 Internal Pump CalibrationReplacing Battery Pack PCBReplacing Sensors Remove shorting clipInstrument Information Enmet Part No DescriptionTarget Specifications Element SpecificationDefault Alarm Default Alarm and ConfigurationsDefault Configuration Target Menu Quick Reference Flow DiagramsBasic Operation Menu Enter any calibration procedure from above Advanced Operation Menu BRH Advanced Maintenance Menu, Software 6.0 version Troubleshooting Symptom Possible Cause ActionWarranty Returning an Instrument for Repair PO Box Fairfield Court Ann Arbor, Michigan 734.761.1270 FaxRepair Return Form