Net Safety Monitoring Inc.
SELECT DISPLAY LANGUAGE
Step 1: After the Restricted Menu the option Select Display Language
YES? displays. Press the Calibration Button or use the Reed Switch to select.
The flashing YES confirms the selection.
Step 2: When the required language displays (English , Espanol , Francais ) press the Calibration Button or use the Reed Switch to select.
The flashing YES confirms the selection.
Note: Factory default is English.
ALARMS
Sensor Fault
WARNING:
The fault detection circuitry does not monitor the operation of external response equipment or the external wiring to these devices. It is important that these devices be checked periodically to ensure they are operational.
Non-standard O2 Levels
When levels >20.9 or <20.9 are detected, the display shows the oxygen level, in percent, on the left side of the display and the LED will blip green. Current output will vary, depending on the O2 level.
If an alarm
Over range is reached if the O2 level exceeds 25%. The display shows <25%O2, the LED goes solid red and 20 mA is delivered.
Sensor Life
Depending on various factors, sensor response may slowly deteriorate over time. If calibration becomes impossible for any reason, the display will show Fail Cal , the analog output will switch repeatedly between 3.0 mA and 3.3 mA and the Status LED alternates Red and Green flashes.
RESET
Manual Reset
A Manual Reset is required after a calibration failure or to clear a latched relay alarm. Simply place and hold the magnet against the Reed Switch or press and hold the Calibration Button for
Remote Reset
If Relay Option is set to Latching (refer to "Setting Relay Options" on page 15) and an open Push Button Switch is connected between the RST terminal and the COM terminal on the Terminal Board, Remote Reset is possible. Also refer to "Remote Reset" on page 9 for details.
17 |