Net Safety Monitoring Inc

Magnetic Reed Switch

The Magnetic Reed Switch is provided to avoid opening the Housing in an environment where gas may be present. The Magnetic Reed Switch functions in the same manner as the Calibration Button but in a non-intrusive manner. Refer to Figure 12, "Controller Functionality", on page 10 for more information.

Attached to the side of the Controller is a magnet:

Place and hold the magnet to the Controller’s Housing (10 o’clock position) to Calibrate and access Main Menu.

Briefly place the magnet to the Controller’s Housing (10 o’clock position) to make a selection (select YES? ).

Power Up

Turn power switch On. A 90 second warm-up routine will begin. The display reads Start Delay Millennium Net Safety , Status LED is Red Slow Flash and current output displays 3.0 mA.

When power is applied, the system is automatically tested to ensure proper functionality.

After warm-up, the Controller will enter normal operation, the display reads %O2, Status LED will Green Blip/blink and analog output will change to 4.0 mA.

Current Loop Measurement (Test Jacks)

Use a standard meter to measure current loop during various states. The Controller’s Housing cover must be removed to access the Test Jacks.

Refer to Table 2, “Status LEDs, Display Messages and Current Loop,” on page 11 for a detailed list and Figure 12, "Controller Functionality", on page 10 for more information.

Status LED

The Status LED will remain solid, flash, blip and/or blink, in either Red or Green, to indicate various states. Refer to Table 2, “Status LEDs, Display Messages and Current Loop,” on page 11 for a detailed list.

THE MAIN MENU

The Main Menu provides access to various functional settings and viewing of current settings.

Calibrate Sensor - Calibrate sensor.

Review Relay Settings - Review the current Fault, Low and High alarm settings. This is a read only section.

Set Relay Options - Set the Low and High alarm settings.

Select Modbus Options.

Restricted Menu - Select low power options.

Select a display language - English, Spanish or French.

Note: The current output will drop to 3.0 mA while in the Main Menu.

Accessing the Main Menu

There are two ways to activate the Main Menu:

Calibration Button found on the faceplate (the Housing cover must be removed to access)

Magnetic Reed Switch (a magnet must be used to activate)

Main Menu Functionality

1.Ensure that the Controller has been turned on and no fault is present.

2.Hold the magnet against the Reed Switch or press and hold the Calibration Button until the message Switch On displays and the countdown (10 to 0) finishes.

3.An option will scroll across the display followed by the prompt YES? .

4.To select an option, momentarily place the magnet to the Reed Switch or press the Calibration Button at the YES? prompt.

5.If you do not wish to select that option wait until the next option appears and then select YES? .

6.A selection is acknowledged with a flashing YES .

7.If no option is selected, the Controller returns to 0 (normal operation).

MLP-A/AR/AD-ST1400-25

12

Image 17
Contents Millennium Page Important Information WarrantyContact Information Table of Contents Maintain How to Return EquipmentController SensorManual ProductOption PlanLocate CONTROLLER/SENSOR External Equipment InstallUnpack Connecting Wires WireField Installation MountSolid State Relay Board Board AssemblyElectromechancial Relay Board Modbus TerminalsTerminal Board Power Controller Sensor Board Terminals Controller WireRS-485 Communication Sensor and ControllerSensor Separation Wiring-Sensor SeparationCurrent Output NON-ISOLATED/ISOLATED WiringTerminal Connection Modbus Termination Remote ResetScrolling 8-character display OperateStatus LED State Current O/P RTU RegistersCalibration Button Power Up Magnetic Reed SwitchAccessing the Main Menu Current Loop Measurement Test JacksCalibrate Calibration ProcedureDisplay ShowsCalibration Failure Review Relay SettingsMonitor Alarm Levels Relay Coil Status Millennium Setting Relay OptionsModel Number Low High Fault Relay OptionsNode Address Set Low Power Options OptionalBaud Rate Modbus OptionsManual Reset Sensor FaultRemote Reset Non-standard O2 LevelsRelays Optional MaintainModbus CurrentDescription Net Safety Part Number Spare PARTS/ACCESSORIEST1Y 7J7 Face Rotation OptionAppendix a Rotate PCB AssemblyAppendix B Resistance Table in Ohms Sensor Specifications Sensor Element