9.0 SPECIFICATIONS
HYDROGEN SULFIDE SPECIFICATIONS
Range:
0 to 100 ppm (for H2S 100)
0 to 500 ppm (for H2S 500) Response Time:
T90 < 30 seconds Operating Temperature Range:
Graphic Display Temperature Range:
(+14 to +140 o F)
Operating Humidity:
15 to 90% RH,
0to 20 oC (32 to 68 oF)
Sensor Accuracy:
±3% of Applied Gas
Linearity:
±3% of Full Scale
Battery Life:
About 100 Hours -
Sensor Cross Sensitivity:
Carbon Monoxide at 300 ppm: < 2 ppm
Sulfur Dioxide, 5 ppm: ≈ - 1 ppm
Nitric Oxide, 35 ppm: < 0.7 ppm
Hydrogen, 10000 ppm: < 10 ppm
Nitrogen Dioxide, 5 ppm: ≈ - 1 ppm
Consult Factory for additional information.
39