5.0 SWITCH FUNCTIONS AND DISPLAYS

All operations on the MINI-SA Portable Detector are controlled by two push- buttons located on the bottom of the unit. The operations vary depending on the operating mode and the graphic display appearing above the two switches. The unit sounds a brief audible “beep” whenever either pushbutton is de- pressed while the instrument is on. Whenever the instrument is turned ON or OFF, three successive audible tones will sound and the Red Alarm LED will flash. The Green Status LED will flash once every fifteen (15) seconds during instrument operation. When in the BATTERY SAVER mode, the Green Sta- tus LED flash rate increases to once every five (5) seconds.

SENSOR

LED

 

(GREEN OR RED

GRAPHIC

DEPENDING

 

PUSHBUTTONS (2)

FIGURE 3

COMPONENT LOCATIONS

When the unit is in the operating mode, depressing the Display Backlight Pushbutton will illuminate the graphic display for fifteen (15) seconds. A light bulb icon for the Backlight is located above the pushbutton on the operat- ing mode screen. Depressing the button again extinguishes the Backlight. The Backlight is sustained ON or OFF while the MENU operations are in progress, except for calibration. In an alarm condition, the Backlight flashes for the duration of the alarm indication.

When in the BATTERY SAVER mode, the instrument graphic display will be blank. Depressing either pushbutton will turn on the graphic display. The display remains on for thirty (30) seconds after the last pushbutton opera- tion.

WARNING

USE OF THE DISPLAY BACKLIGHT FOR EXTENDED PERIODS OF TIME WILL REDUCE BATTERY SERVICE LIFE.

10

Image 10 Contents
MINI-SA Portable GAS Detection Instrument 39NO US Scott Instruments Scott Instruments MINI-SA Portable GAS Detector Introduction Operational EnvironmentRegistration Battery Installation General Instructions UnpackingInstrument Description H2S Sensor shown for referenceAlarm Legends Audible Alarms Vibration Alarm Status LightGraphic Switch Functions and DisplaysSensor Pushbuttons Component LocationsOperating Instructions MINI-SA Version Instrument Turn on / OFFScott Instruments MINI-SA Version O2 Sensor InitializingMenu Do YOU Want to Turn Unit OFF? YesnoMenu Selection Peak / Hold Next Select Nextselect O2 Alarm SetpointNextselect Configuration Nextselect Zero Calibration Menu Auto Zero Required Before Span Calibration Acknowledge Span Calibration Menu for Toxic GAS onlyChange SET DownExit To GAS Cylinder Calibration AdapterSpan Calibration Approaching Failure Sensor Replacement MessageMent Message Replace SoonReset Show Peak Select Mode Peak HoldReset Show Peak Peak O2 Level Peak / Hold MenuReset Hide Peak Peak O2 Level Reset Hide PeakReset Hold Mode Activate ExitHold Activated Alarm Setpoint Menu Change Alarm SetpointChange O2 Alarm Down SETChange O2 Alarm Setpoint NON-LATCHING Latchok NON-LATCHING Latchok O2 AlarmTWA / Stel Menu Toxic Gases only Resetdisable Reset Show Disable TWA/STELStel TWA Resetenable Reset Show ON/OFF TWA/STELConfiguration Battery Password Saver Menu Configuration MenuBattery Saver Mode Configuration Enable Battery Saver MODE? YesnoPassword Protection Disable Battery Saver MODE? YesnoEnter Password Upset Password Protection For Advanced Menus Turnchange Onpassword Password Required For Advanced Menus Enter Exit PasswordEnter NEW Password Upset Verify NEW Password Upset Turn OFF Recal Oxygen Sensor Turn OFF RecalCalibration Gases Specifications Hydrogen Sulfide Specifications Carbon Monoxide Specifications Sulphur Dioxide Specification Phosphine Specifications Chlorine Specification Ammonia Specifications Chlorine Dioxide Specification Oxygen Specifications Hydrogen Cyanide Specification Nitrogen Dioxide Specifications