Fluke Rev. 4 user manual Measuring Electrical Parameters

Page 38

701/702

Users Manual

If the range is locked, overrange inputs produce a display of - - - - - -. In Auto Range, out of range inputs produce a display of ! ! ! ! ! !.

Measuring Electrical Parameters

When you turn on the calibrator, it powers up in the dc voltage measurement function. To select an electrical measurement function from either SOURCE or MEASURE/SOURCE mode, first press

Mfor MEASURE mode, then proceed as follows:

1.Press Mfor current, vfor dc voltage, h once for ac voltage or twice for frequency, or qfor resistance.

NOTE

To measure frequency you are prompted to select a frequency range as shown next. If you expect the frequency you are measuring to be below 20 Hz, press dto select the lower frequency range, then press e.

2.Connect the test leads as shown in Figure 9, depending on the measurement function.

28

Image 38
Contents 701/702 Page Table of Contents 701/702 Iii 100 List of Tables 701/702 List of Figures 103 Documenting Process Calibrator Introduction701/702 Source FunctionMeasure Standard Equipment Standard Equipment Users ManualMail-in Card Documenting Process Calibrator Accessories PMLink Software Safety Information CATSafety Information Calibrator FeatureGetting Acquainted with Your Input and Output JacksGetting Acquainted with Your Calibrator Keys KeysKey Functions Key Numeric keypad Display Setting Up the CalibratorUsing the Strap and Bail Elements of a Typical Display Installing the Strap and Using the Bail Charging the Battery Battery Life Removing the Battery and Using the ChargerBacklight on Operating ModesBacklight OFF Automatically Saving Battery Life Using the Optional Battery EliminatorSelecting the Display Language Adjusting the Display ContrastSetting Up the Date and Time Display Using the Backlight Personalizing the Calibrator Using Measure Mode Measurement RangesMeasuring Electrical Parameters Electrical Measurement Connections Testing Continuity Measuring PressureDocumenting Process Calibrator Connections for Measuring Pressure Measuring Temperature Using ThermocouplesPositive Lead Color ANSI* IEC TypePositive Lead Material Negative Lead MaterialMeasuring Temperature with a Thermocouple RTD Type ICE Point R0 Material Range CDocumenting Process Calibrator Wrong Measuring Temperature with an RTD Measurements in Percent of Scale Damping Your MeasurementsUsing Source Mode Sourcing Electrical ParametersElectrical Sourcing Connections Simulating a 4 to 20 mA Transmitter Connections for Simulating a 4 to 20 mA Transmitter Providing Loop Power Connections for Supplying Loop Power Sourcing Pressure 701/702 Connections for Sourcing Pressure Simulating Thermocouples Connections for Simulating a Thermocouple Simulating RTDs PS S +PS Z +IN Stepping the Output Sourcing in Percent of ScaleStepping and Ramping the Output Value Ramping the Output 701/702 Documenting Process Calibrator Checking a Relay Output Trip Alarm Simultaneous Measure/Source Measure Function Source Function 2W RTDSource Function Calibrating a Process Instrument Generating As Found Test DataCalibrating a Process Instrument 701/702 Documenting Process Calibrator Adjusting the Transmitter As Left Test Run Memory Operations Saving ResultsReviewing Memory Memory OperationsData Logging Model 702 Only Reading Rate MAX. Number Observing Min and Max Measurements Viewing the Task List Model 702 OnlyQuick Guide to Applications Clearing MemoryCalibrating a Chart Recorder Monitoring AC Line Voltage and Frequency Calibrating a Current-to-Pressure I/P Transducer Measuring the Output Current of a Transmitter Measuring a Precision Resistor Checking a Switch Calibrating a Pressure-to-Current P/I Transducer Calibrating a mV to Current Transmitter Checking a Vortex Shedding Flowmeter Internal Lithium Backup Battery MaintenanceReplacing the Battery Pack Cleaning the CalibratorCalibration Data Service Center Calibration or RepairReplacement Parts Maintenance Specifications Range ResolutionDC Voltage Measurement Model Reading +% of Full Scale YearReading + Number of Counts Year AC Voltage MeasurementFrequency Range DC Current Measurement Range Resolution Model Reading +% of Full ScaleRange Resolution Model Resistance MeasurementContinuity Testing Frequency MeasurementModel Output +% of Full Scale Output + % of Full Scale Range ResolutionsDC Voltage Output DC Current Output Output +% of Full Scale YearRanges Resolution Model Output + Ohms Year Resistance SourcingFrequency Sourcing Frequency RangesMeasure Degrees C Year Source Degrees C Temperature, ThermocouplesTypes C Temperture, Thermocouples Maximum Input Voltage 30 Temperature, Resistance Temperature Detectors Measure Degrees C Source Degrees C YearLoop Power Supply SettingTop and Bottom Limits of Ranges with Auto Range On RANGE, Frequency Measure RANGE, Measure V CCRANGE, Current Source General Specifications SafetyOperating Environment Specification 104 Index 701/702 Index