701/702
Users Manual
If the range is locked, overrange inputs produce a display of - - - - -
Measuring Electrical Parameters
When you turn on the calibrator, it powers up in the dc voltage measurement function. To select an electrical measurement function from either SOURCE or MEASURE/SOURCE mode, first press
Mfor MEASURE mode, then proceed as follows:
1.Press Mfor current, vfor dc voltage, h once for ac voltage or twice for frequency, or qfor resistance.
NOTE
To measure frequency you are prompted to select a frequency range as shown next. If you expect the frequency you are measuring to be below 20 Hz, press dto select the lower frequency range, then press e.
2.Connect the test leads as shown in Figure 9, depending on the measurement function.
28