701/702
Users Manual
Measurements in Percent of Scale
To display measurements as a percent of scale as shown below, proceed as follows:
1.If necessary, press Mfor MEASURE mode.
2.Select a measurement function (m, v, q, t, or p) as previously described.
3.Press the Scale softkey.
4.Use the numeric keypad to enter the 0% of scale value.
5.Press e.
6.Use the numeric keypad to enter the 100% of scale value.
7.Press e.
8.Press the Done softkey.
Percent of scale remains in effect until you press the Scale softkey again or change to another measurement function.
Damping Your Measurements
The calibrator normally applies a software filter to dampen measurements in all functions except continuity. The specifications assume that damping is turned on. The damping method is a running average of the last several measurements. Fluke recommends that you leave damping on. Turning damping off may be useful when measurement response is more important than accuracy or noise reduction. If you want to turn off damping, press the More Choices softkey twice, then press the Dampen softkey so that Off appears. Press Dampen again to turn damping back on. The default state is On.
40