Fluke Rev. 4 Type, Positive Lead Material, Positive Lead Color ANSI* IEC, Negative Lead Material

Page 44

701/702

Users Manual

Table 5. Thermocouple Types Accepted

TYPE

E

N

J

K

T

B

R

S

C***

L (DIN J) U (DIN T)

POSITIVE LEAD

MATERIAL

Chromel

Ni-Cr-Si

Iron

Chromel

Copper

Platinum (30% Rhodium)

Platinum (13% Rhodium)

Platinum (10% Rhodium)

Tungsten (5% Rhenium)

Iron

Copper

POSITIVE LEAD

(H)COLOR

ANSI* IEC**

Purple

Violet

 

 

Orange

 

 

 

White

Black

 

 

Yellow

Green

 

 

Blue

Brown

 

 

Gray

 

 

 

Black

Orange

 

 

Black

Orange

 

 

White

 

 

 

 

 

 

 

NEGATIVE LEAD

MATERIAL

Constantan

Ni-Si-Mg

Constantan

Alumel

Constantan

Platinum (6% Rhodium)

Platinum

Platinum

Tungsten (26% Rhenium)

Constantan

Constantan

SPECIFIED RANGE

(°C)

-250 to 1000

-270 to 1300

-210 to 1200

-270 to 1372

-250 to 400

600 to 1820

-20 to 1767

-20 to 1767

0 to 2316

-200 to 900

-200 to 600

*American National Standards Institute (ANSI) device negative lead (L) is always red. **International Electrotechnical Commission (IEC) device negative lead (L) is always white.

*** Not an ANSI designation but a Hoskins Engineering Company designation.

34

Image 44
Contents 701/702 Page Table of Contents 701/702 Iii 100 List of Tables 701/702 List of Figures 103 Documenting Process Calibrator Introduction701/702 Source FunctionMeasure Standard Equipment Standard Equipment Users ManualMail-in Card Documenting Process Calibrator Accessories PMLink Software Safety Information CATSafety Information Feature Getting Acquainted with YourCalibrator Input and Output JacksGetting Acquainted with Your Calibrator Keys KeysKey Functions Key Numeric keypad Display Setting Up the CalibratorUsing the Strap and Bail Elements of a Typical Display Installing the Strap and Using the Bail Charging the Battery Battery Life Removing the Battery and Using the ChargerBacklight on Operating ModesBacklight OFF Automatically Saving Battery Life Using the Optional Battery EliminatorSelecting the Display Language Adjusting the Display ContrastSetting Up the Date and Time Display Using the Backlight Personalizing the Calibrator Using Measure Mode Measurement RangesMeasuring Electrical Parameters Electrical Measurement Connections Testing Continuity Measuring PressureDocumenting Process Calibrator Connections for Measuring Pressure Measuring Temperature Using ThermocouplesType Positive Lead MaterialPositive Lead Color ANSI* IEC Negative Lead MaterialMeasuring Temperature with a Thermocouple RTD Type ICE Point R0 Material Range CDocumenting Process Calibrator Wrong Measuring Temperature with an RTD Measurements in Percent of Scale Damping Your MeasurementsUsing Source Mode Sourcing Electrical ParametersElectrical Sourcing Connections Simulating a 4 to 20 mA Transmitter Connections for Simulating a 4 to 20 mA Transmitter Providing Loop Power Connections for Supplying Loop Power Sourcing Pressure 701/702 Connections for Sourcing Pressure Simulating Thermocouples Connections for Simulating a Thermocouple Simulating RTDs PS S +PS Z +IN Stepping the Output Sourcing in Percent of ScaleStepping and Ramping the Output Value Ramping the Output 701/702 Documenting Process Calibrator Checking a Relay Output Trip Alarm Simultaneous Measure/Source Measure Function Source Function 2W RTDSource Function Calibrating a Process Instrument Generating As Found Test DataCalibrating a Process Instrument 701/702 Documenting Process Calibrator Adjusting the Transmitter As Left Test Run Memory Operations Saving ResultsReviewing Memory Memory OperationsData Logging Model 702 Only Reading Rate MAX. Number Observing Min and Max Measurements Viewing the Task List Model 702 OnlyQuick Guide to Applications Clearing MemoryCalibrating a Chart Recorder Monitoring AC Line Voltage and Frequency Calibrating a Current-to-Pressure I/P Transducer Measuring the Output Current of a Transmitter Measuring a Precision Resistor Checking a Switch Calibrating a Pressure-to-Current P/I Transducer Calibrating a mV to Current Transmitter Checking a Vortex Shedding Flowmeter Maintenance Replacing the Battery PackInternal Lithium Backup Battery Cleaning the CalibratorCalibration Data Service Center Calibration or RepairReplacement Parts Maintenance Specifications Resolution DC Voltage MeasurementRange Model Reading +% of Full Scale YearReading + Number of Counts Year AC Voltage MeasurementFrequency Range Range Resolution Model Reading +% of Full Scale Range Resolution ModelDC Current Measurement Resistance MeasurementContinuity Testing Frequency MeasurementModel Output +% of Full Scale Output + % of Full Scale Range ResolutionsDC Voltage Output DC Current Output Output +% of Full Scale YearRanges Resolution Model Output + Ohms Year Resistance SourcingFrequency Sourcing Frequency RangesMeasure Degrees C Year Source Degrees C Temperature, ThermocouplesTypes C Temperture, Thermocouples Maximum Input Voltage 30 Temperature, Resistance Temperature Detectors Measure Degrees C Source Degrees C YearLoop Power Supply SettingTop and Bottom Limits of Ranges with Auto Range On RANGE, Frequency Measure RANGE, Measure V CCRANGE, Current Source General Specifications SafetyOperating Environment Specification 104 Index 701/702 Index