Documenting Process Calibrator
Specifications
DC Voltage Measurement
RANGE
110mV
1.1 V
RESOLUTION
1μV
10μV
MODEL 702
% OF READING +% OF FULL SCALE
1 YEAR | 2 YEAR |
0.025% + 0.015% | 0.05% + 0.015% |
0.025% + 0.005% | 0.05% + 0.005% |
MODEL 701
% OF READING +% OF FULL SCALE
1 YEAR | 2 YEAR |
0.05% + 0.015% | 0.1% + 0.015% |
0.05% + 0.005% | 0.1% + 0.005% |
11 V | 100 μV | 0.025% + 0.005% | 0.05% + 0.005% | 0.05% + 0.005% | 0.1% + 0.005% |
110 V | 1 mV | 0.05% + 0.005% | 0.1% + 0.005% | 0.05% + 0.005% | 0.1% + 0.005% |
300 V | 10 mV | 0.05% + 0.005% | 0.1% + 0.005% | 0.05% + 0.005% | 0.1% + 0.005% |
Temperature Coefficient: (0.001% of rdg. + 0.0015% f.s.)/°C in the ranges
Input Impedance: 5 MΩ
Common Mode Error: 0.008% f.s./(Common Mode Volt)
Maximum Input Voltage: 300V rms
87