Fluke Rev. 4 user manual Function, Measure, Source

Page 13

Documenting Process Calibrator

Introduction

Table 1. Summary of Source and Measure Functions

FUNCTION

vdc V h ac V h Frequency qResistance mdc Current qContinuity

MEASURE

0 to ±300V

0 to 300V rms, 20 Hz to 5 kHz

1 Hz to 50 kHz, 110 mV to 300V rms

0Ω to 11 kΩ

0 to 110 mA

Beep and the word Short indicates continuity

SOURCE

0 to 11V in V or mV units (10 mA max)

No sourcing

1 mV to 10V pk square wave, 2 Hz to 50 kHz

0Ω to 11 kΩ

0 to 22 mA (28V max)

No sourcing

tThermocouple

tRTD

Measure 2,3, or 4 wire.

Source 2 Wire.

E, N, J, L, K, T, U, B, R, S, or C

100Ω Platinum (3926)

100Ω Platinum (385)

120Ω Nickel (672)

200Ω Platinum (385)

500Ω Platinum (385)

1000Ω Platinum (385)

10Ω Copper (427)

100Ω Platinum (3916)

pPressure

14 modules ranging from 0 to 10” H20 (2.5 kPa) to 0 to 10,000 psi (69,000 kPa)

Note

sLoop Power

24 or 28V

(22 mA max)

Note: Use an external hand pump or other pressure source as a pressure

stimulus for the source pressure function.

3

Image 13
Contents 701/702 Page Table of Contents 701/702 Iii 100 List of Tables 701/702 List of Figures 103 Introduction Documenting Process Calibrator701/702 Measure FunctionSource Standard Equipment Standard Equipment Users ManualMail-in Card Documenting Process Calibrator Accessories PMLink Software CAT Safety InformationSafety Information Getting Acquainted with Your FeatureCalibrator Input and Output JacksGetting Acquainted with Your Calibrator Keys KeysKey Functions Key Numeric keypad Using the Strap and Bail Setting Up the CalibratorDisplay Elements of a Typical Display Installing the Strap and Using the Bail Charging the Battery Removing the Battery and Using the Charger Battery LifeBacklight OFF Operating ModesBacklight on Using the Optional Battery Eliminator Automatically Saving Battery LifeSetting Up the Date and Time Display Adjusting the Display ContrastSelecting the Display Language Using the Backlight Personalizing the Calibrator Measurement Ranges Using Measure ModeMeasuring Electrical Parameters Electrical Measurement Connections Measuring Pressure Testing ContinuityDocumenting Process Calibrator Connections for Measuring Pressure Using Thermocouples Measuring TemperaturePositive Lead Material TypePositive Lead Color ANSI* IEC Negative Lead MaterialMeasuring Temperature with a Thermocouple Range C RTD Type ICE Point R0 MaterialDocumenting Process Calibrator Wrong Measuring Temperature with an RTD Damping Your Measurements Measurements in Percent of ScaleSourcing Electrical Parameters Using Source ModeElectrical Sourcing Connections Simulating a 4 to 20 mA Transmitter Connections for Simulating a 4 to 20 mA Transmitter Providing Loop Power Connections for Supplying Loop Power Sourcing Pressure 701/702 Connections for Sourcing Pressure Simulating Thermocouples Connections for Simulating a Thermocouple Simulating RTDs PS S +PS Z +IN Stepping and Ramping the Output Value Sourcing in Percent of ScaleStepping the Output Ramping the Output 701/702 Documenting Process Calibrator Checking a Relay Output Trip Alarm Simultaneous Measure/Source Source Function 2W RTD Measure FunctionSource Function Generating As Found Test Data Calibrating a Process InstrumentCalibrating a Process Instrument 701/702 Documenting Process Calibrator Adjusting the Transmitter As Left Test Run Saving Results Memory OperationsMemory Operations Reviewing MemoryData Logging Model 702 Only Reading Rate MAX. Number Viewing the Task List Model 702 Only Observing Min and Max MeasurementsClearing Memory Quick Guide to ApplicationsCalibrating a Chart Recorder Monitoring AC Line Voltage and Frequency Calibrating a Current-to-Pressure I/P Transducer Measuring the Output Current of a Transmitter Measuring a Precision Resistor Checking a Switch Calibrating a Pressure-to-Current P/I Transducer Calibrating a mV to Current Transmitter Checking a Vortex Shedding Flowmeter Replacing the Battery Pack MaintenanceInternal Lithium Backup Battery Cleaning the CalibratorReplacement Parts Service Center Calibration or RepairCalibration Data Maintenance Specifications DC Voltage Measurement ResolutionRange Model Reading +% of Full Scale YearFrequency Range AC Voltage MeasurementReading + Number of Counts Year Range Resolution Model Range Resolution Model Reading +% of Full ScaleDC Current Measurement Resistance MeasurementFrequency Measurement Continuity TestingDC Voltage Output Range ResolutionsModel Output +% of Full Scale Output + % of Full Scale Output +% of Full Scale Year DC Current OutputResistance Sourcing Ranges Resolution Model Output + Ohms YearFrequency Ranges Frequency SourcingTypes C Temperature, ThermocouplesMeasure Degrees C Year Source Degrees C Temperture, Thermocouples Maximum Input Voltage 30 Measure Degrees C Source Degrees C Year Temperature, Resistance Temperature DetectorsSetting Loop Power SupplyTop and Bottom Limits of Ranges with Auto Range On RANGE, Current Source RANGE, Measure V CCRANGE, Frequency Measure Safety General SpecificationsOperating Environment Specification 104 Index 701/702 Index