701/702
Users Manual
Simulating Thermocouples
NOTE
Refer to “Measuring Temperature” earlier in the manual for a table of data relating to thermocouple types supported by the calibrator.
Connect the calibrator TC connector to the instrument under test with thermocouple wireand theappropriate thermocouple
1.Attach the thermocouple leads to the appropriate TC miniplug, then to the TC jack as Figure 12 shows.
2.If necessary, press Mfor SOURCE mode.
3.Press tfor the display that prompts you to enter thermocouple type.
4.Press the uor dkey followed by eto select the desired thermocouple type.
5.Enter the temperature you want to simulate as prompted by the display and press e.
NOTE
If you use copper wire instead of thermocouple wire, the reference junction is no longer inside the calibrator. The reference junction is moved to the instrument (transmitter, indicator, controller, etc.) input terminals. You must measure this external reference temperature accurately and enter it into the calibrator. Do this by pressing sand setting Ref. Junc. Compensat and Ref. Junc. Temp. After you enter the external reference temperature, the calibrator corrects all voltages to compensate for this new reference junction temperature.