To avoid damaging the pressure module from corrosion, use them only with specified materials. Refer to the pressure module instruction sheet for the specified material compatibility for your pressure module.

1.Connect a pressure module and pressure source to the calibrator as Figure 17 shows. The threads on the pressure modules accept ¼ NPT fittings.

Use the supplied ¼ NPT to ¼ ISO adapter if necessary.

2.If necessary, press Mfor SOURCE mode.

3.Press p. The calibrator automatically senses which pressure module is attached and sets its range accordingly.

4.To zero the pressure module, vent the module to atmosphere, and press c. Repeat this step until the reading is zeroed. You must perform this step before you execute a PMLink Task that sources pressure

5.Pressurize the pressure line with the pressure source to the desired level as shown on the display.

6.If necessary, you can change pressure display

units to psi, mmHg, inHg, inH2O, ftH2O, bar, or kPa as follows:

a.Press s

b.Press Next Page twice.

c.Press ewith the cursor on Pressure Units.

d.Select the pressure units with the uor d keys.

e.Press e.

f.Press the Done softkey.

48

Image 58
Contents 701/702 Page Table of Contents 701/702 Iii 100 List of Tables 701/702 List of Figures 103 Documenting Process Calibrator Introduction701/702 Measure FunctionSource Standard Equipment Standard Equipment Users ManualMail-in Card Documenting Process Calibrator Accessories PMLink Software Safety Information CATSafety Information Calibrator FeatureGetting Acquainted with Your Input and Output JacksGetting Acquainted with Your Calibrator Keys KeysKey Functions Key Numeric keypad Using the Strap and Bail Setting Up the CalibratorDisplay Elements of a Typical Display Installing the Strap and Using the Bail Charging the Battery Battery Life Removing the Battery and Using the ChargerBacklight OFF Operating ModesBacklight on Automatically Saving Battery Life Using the Optional Battery EliminatorSetting Up the Date and Time Display Adjusting the Display ContrastSelecting the Display Language Using the Backlight Personalizing the Calibrator Using Measure Mode Measurement RangesMeasuring Electrical Parameters Electrical Measurement Connections Testing Continuity Measuring PressureDocumenting Process Calibrator Connections for Measuring Pressure Measuring Temperature Using ThermocouplesPositive Lead Color ANSI* IEC TypePositive Lead Material Negative Lead MaterialMeasuring Temperature with a Thermocouple RTD Type ICE Point R0 Material Range CDocumenting Process Calibrator Wrong Measuring Temperature with an RTD Measurements in Percent of Scale Damping Your MeasurementsUsing Source Mode Sourcing Electrical ParametersElectrical Sourcing Connections Simulating a 4 to 20 mA Transmitter Connections for Simulating a 4 to 20 mA Transmitter Providing Loop Power Connections for Supplying Loop Power Sourcing Pressure 701/702 Connections for Sourcing Pressure Simulating Thermocouples Connections for Simulating a Thermocouple Simulating RTDs PS S +PS Z +IN Stepping and Ramping the Output Value Sourcing in Percent of ScaleStepping the Output Ramping the Output 701/702 Documenting Process Calibrator Checking a Relay Output Trip Alarm Simultaneous Measure/Source Measure Function Source Function 2W RTDSource Function Calibrating a Process Instrument Generating As Found Test DataCalibrating a Process Instrument 701/702 Documenting Process Calibrator Adjusting the Transmitter As Left Test Run Memory Operations Saving ResultsReviewing Memory Memory OperationsData Logging Model 702 Only Reading Rate MAX. Number Observing Min and Max Measurements Viewing the Task List Model 702 OnlyQuick Guide to Applications Clearing MemoryCalibrating a Chart Recorder Monitoring AC Line Voltage and Frequency Calibrating a Current-to-Pressure I/P Transducer Measuring the Output Current of a Transmitter Measuring a Precision Resistor Checking a Switch Calibrating a Pressure-to-Current P/I Transducer Calibrating a mV to Current Transmitter Checking a Vortex Shedding Flowmeter Internal Lithium Backup Battery MaintenanceReplacing the Battery Pack Cleaning the CalibratorReplacement Parts Service Center Calibration or RepairCalibration Data Maintenance Specifications Range ResolutionDC Voltage Measurement Model Reading +% of Full Scale YearFrequency Range AC Voltage MeasurementReading + Number of Counts Year DC Current Measurement Range Resolution Model Reading +% of Full ScaleRange Resolution Model Resistance MeasurementContinuity Testing Frequency MeasurementDC Voltage Output Range ResolutionsModel Output +% of Full Scale Output + % of Full Scale DC Current Output Output +% of Full Scale YearRanges Resolution Model Output + Ohms Year Resistance SourcingFrequency Sourcing Frequency RangesTypes C Temperature, ThermocouplesMeasure Degrees C Year Source Degrees C Temperture, Thermocouples Maximum Input Voltage 30 Temperature, Resistance Temperature Detectors Measure Degrees C Source Degrees C YearLoop Power Supply SettingTop and Bottom Limits of Ranges with Auto Range On RANGE, Current Source RANGE, Measure V CCRANGE, Frequency Measure General Specifications SafetyOperating Environment Specification 104 Index 701/702 Index