Fluke Rev. 4 user manual Calibrating a mV to Current Transmitter

Page 91

Documenting Process Calibrator

Quick Guide to Applications

701 DOCUMENTING PROCESS CALIBRATOR

Red

 

MEAS

mA

SETUP

 

 

SOURCE

 

 

 

 

 

V

V

 

TC

 

Hz

 

RTD

 

 

 

 

7

8

9

CLEAR

RANGE

( ZERO)

4

5

6

 

 

1

2

3

 

 

 

0

.

ENTER

V

mA

mA

V

 

RTD

 

RTD

 

 

 

 

 

 

 

SOURCE

MEAS

300V

TC

 

MAX

 

30V MAX

 

 

 

Measure mA

Input

0 – 1mV, 0 – 10mV, 0– 100mV, 1 – 5V, 0 – 1V, 0 – 10V

Output 4–20 mA

Black

Red

Source mV or V

Output

+

Input

+

Black

Figure 32. Calibrating a mV to Current Transmitter

81

Image 91
Contents 701/702 Page Table of Contents 701/702 Iii 100 List of Tables 701/702 List of Figures 103 Introduction Documenting Process Calibrator701/702 Measure FunctionSource Standard Equipment Standard Equipment Users ManualMail-in Card Documenting Process Calibrator Accessories PMLink Software CAT Safety InformationSafety Information Input and Output Jacks FeatureGetting Acquainted with Your CalibratorGetting Acquainted with Your Calibrator Keys KeysKey Functions Key Numeric keypad Using the Strap and Bail Setting Up the CalibratorDisplay Elements of a Typical Display Installing the Strap and Using the Bail Charging the Battery Removing the Battery and Using the Charger Battery LifeBacklight OFF Operating ModesBacklight on Using the Optional Battery Eliminator Automatically Saving Battery LifeSetting Up the Date and Time Display Adjusting the Display ContrastSelecting the Display Language Using the Backlight Personalizing the Calibrator Measurement Ranges Using Measure ModeMeasuring Electrical Parameters Electrical Measurement Connections Measuring Pressure Testing ContinuityDocumenting Process Calibrator Connections for Measuring Pressure Using Thermocouples Measuring TemperatureNegative Lead Material TypePositive Lead Material Positive Lead Color ANSI* IECMeasuring Temperature with a Thermocouple Range C RTD Type ICE Point R0 MaterialDocumenting Process Calibrator Wrong Measuring Temperature with an RTD Damping Your Measurements Measurements in Percent of ScaleSourcing Electrical Parameters Using Source ModeElectrical Sourcing Connections Simulating a 4 to 20 mA Transmitter Connections for Simulating a 4 to 20 mA Transmitter Providing Loop Power Connections for Supplying Loop Power Sourcing Pressure 701/702 Connections for Sourcing Pressure Simulating Thermocouples Connections for Simulating a Thermocouple Simulating RTDs PS S +PS Z +IN Stepping and Ramping the Output Value Sourcing in Percent of ScaleStepping the Output Ramping the Output 701/702 Documenting Process Calibrator Checking a Relay Output Trip Alarm Simultaneous Measure/Source Source Function 2W RTD Measure FunctionSource Function Generating As Found Test Data Calibrating a Process InstrumentCalibrating a Process Instrument 701/702 Documenting Process Calibrator Adjusting the Transmitter As Left Test Run Saving Results Memory OperationsMemory Operations Reviewing MemoryData Logging Model 702 Only Reading Rate MAX. Number Viewing the Task List Model 702 Only Observing Min and Max MeasurementsClearing Memory Quick Guide to ApplicationsCalibrating a Chart Recorder Monitoring AC Line Voltage and Frequency Calibrating a Current-to-Pressure I/P Transducer Measuring the Output Current of a Transmitter Measuring a Precision Resistor Checking a Switch Calibrating a Pressure-to-Current P/I Transducer Calibrating a mV to Current Transmitter Checking a Vortex Shedding Flowmeter Cleaning the Calibrator MaintenanceReplacing the Battery Pack Internal Lithium Backup BatteryReplacement Parts Service Center Calibration or RepairCalibration Data Maintenance Specifications Model Reading +% of Full Scale Year ResolutionDC Voltage Measurement RangeFrequency Range AC Voltage MeasurementReading + Number of Counts Year Resistance Measurement Range Resolution Model Reading +% of Full ScaleRange Resolution Model DC Current MeasurementFrequency Measurement Continuity TestingDC Voltage Output Range ResolutionsModel Output +% of Full Scale Output + % of Full Scale Output +% of Full Scale Year DC Current OutputResistance Sourcing Ranges Resolution Model Output + Ohms YearFrequency Ranges Frequency SourcingTypes C Temperature, ThermocouplesMeasure Degrees C Year Source Degrees C Temperture, Thermocouples Maximum Input Voltage 30 Measure Degrees C Source Degrees C Year Temperature, Resistance Temperature DetectorsSetting Loop Power SupplyTop and Bottom Limits of Ranges with Auto Range On RANGE, Current Source RANGE, Measure V CCRANGE, Frequency Measure Safety General SpecificationsOperating Environment Specification 104 Index 701/702 Index