Chapter 1 General Information

Table 1-1.Performance Specifications (1 of 2)

Specifications are valid when the unit is calibrated at ambient temperature after a 5 minute warmup. Typical values are provided for reference only and are not guaranteed.

Description

Site Master:

S113C, S114C

S331C, S332C

Frequency Accuracy (RF Source Mode) Frequency Resolution: S113C, S114C S331C, S332C

SWR: Range Resolution

Return Loss: Range Resolution

**Distance-To-Fault (DTF):

Vertical Range Return Loss: SWR:

Horizontal Range

Horizontal Resolution for Coax (rectangular windowing)

Value

Frequency Range:

2 to 1600 MHz

25 to 4000 MHz

￿75 parts per million @ 25°C*

10kHz

100kHz 1.00 to 65.00 0.01

0.00 to 54.00 dB

0.01 dB

0.00to 54.00 dB

1.00to 65.00

0 to ((# of data points –1) ￿ resolution) a maximum of 1000m (3281 ft.) with a maximum of 517 points resolution, # of data pts. = 130, 259, 517

(1.5 ￿ 108)(Vp)

￿F

Where Vp is the relative propagation velocity of the cable; dp is the number of data points

(130, 259, 517); ￿F is the stop frequency minus the start frequency (Hz)

Horizontal Resolution for Waveguide

 

 

 

 

1.5 ￿108 ( 1￿ (F / F )2)

 

 

 

c

1

 

 

￿ F

 

 

 

Where FC is the waveguide cutoff frequency (Hz);

 

F1 is the start frequency (Hz); ￿F is the stop fre-

 

quency minus the start frequency (Hz)

RF Power Monitor:

 

 

 

 

Display Range

–80.0 to +80 dBm or

 

10.0 pW to 100.0 kW

Detector Range

–50 dBm to +20 dBm, or 10 ￿W to 100 mW

Offset Range

0 to +60.0 dB

 

 

Resolution

0.1 dB or 0.1 W

 

 

Test Port Connector

Type N, 50￿, female

***Immunity to Interfering signals

on-channel

on-frequency

up to the level of: S113C, S114C

+17 dBm

+10 dBm

S331C, S332C

+17 dBm

–6 dBm

Maximum Input without Damage:

 

 

 

 

Test Port, Type N (f)

+20 dBm, 50￿, +50 VDC

RF Power Detector

+20 dBm, 50￿, +50 VDC

Measurement Accuracy:

Measurement accuracy depends on calibration components. Precision calibration components have a directivity of 42 dB.

Cable Insertion Loss:

Range

0.00 to 54.00 dB

 

Resolution

0.01 dB

Transmission Line Loss (one-port)

 

 

Range

0.00 to 20.00 dB

 

Resolution

0.01 dB

1 - 4

Image 11 Contents
Users Guide Warranty Page Table of Contents Power Measurement Iii/iv Page Description Standard AccessoriesChapter General Information IntroductionOptions Optional AccessoriesPerformance Specifications Description Site Master Spectrum Analyzer Calibration Preventive MaintenanceInstaCal Module Annual VerificationAnritsu Service Centers Chapter Functions and Operations Test Connector PanelActive Function Block Soft Keys Front Panel OverviewMode Function Hard KeysClear Keypad Hard KeysAuto MODE=FREQ FREQ/DIST Softkeys Soft KeysMODE=DTF FREQ/DIST Softkeys MODE=SPECTRUM Analyzer FREQ/DIST Amplitude Softkeys Power Monitor Mode Soft Key Labels Functions and Operations Functions and Operations Amplitude RUN/HOLD button Functions and Operations Power Monitor Menu Functions and Operations Limit SYS Self Test SymbolsError Codes Self Test ErrorsError Message ERRORD2 Range ErrorsInstalled InstaCal Error MessagesError MessageDescription Charging the Battery in the Optional Charger Battery InformationCharging a New Battery Charging the Battery in the Site MasterBattery Life Battery Charge IndicatorImportant Battery Information Chapter Getting Started Power On ProcedureCalibration Select the Frequency or DistanceCalibration Verification Trace Characteristics in Return Loss ModeCalibration Manual Calibration ProcedureOffset Method InstaCal Module VerificationTermination Method Calibration with the Test Port Extension Cable InstaCal Module Calibration ProcedureAuto Scale Setting the ScaleSave and Recall a Setup Saving a SetupRecalling a Display Recalling a SetupSave and Recall a Display Saving a DisplayChanging the Display Language Set the Distance and Cable TypeChanging the Units Printing a Screen PrintingPrinter Printer Switch SettingsUsing the Site Master Soft Carrying Case Using the Soft Carrying CaseLine Sweep Fundamentals Chapter Cable & Antenna MeasurementsCable & Antenna Measurements Information Required for a Line SweepSystem Return Loss Measurement Typical Line Sweep Test ProceduresInsertion Loss Measurement Procedure Cable Loss ModeSet M1 to Marker to Peak Set M2 to Marker to Valley Typical Transmission Line Cable Loss Trace Procedure Return Loss ModeTypical Return Loss Trace Distance-To-Fault DTF Transmission Line Test Shows an example of a typical DTF return loss trace Procedure DTF-SWR ModeAntenna Subsystem Return Loss Test 11/4-12 Page Occupied Bandwidth Chapter Spectrum Analyzer MeasurementsChannel Power Measurement with the Site Master Channel Power MeasurementProcedure Example Acpr Specification Adjacent Channel Power MeasurementCW Wave Out-of-Band Spurious Emission Measurements In-Band/Out-of-Channel Measurements Field Strength Measurement VdBV PdBm Antenna CalculationsPage Power Measurement Chapter Power MeasurementDisplaying Relative Power System Requirements FeaturesChapter Site Master Software Tools Communication Port Setting InstallationDownloading Traces from the Site Master Using Software ToolsInterface Cable Installation Graph Titles Display ModePlot Capture Plot PropertiesMisc Trace Overlay or Plot OverlaySaving Traces MarkersCreating a Custom Cable List Custom Cable ListSaving a Plot to a Spreadsheet Querying the Site Master Cable List Entering Antenna FactorsUploading a Cable List Printing Formats Uploading Antenna FactorsCreating a Database Appendix a Reference Data Coaxial Cable Technical DataAppendix a Reference Data A-4 Page Examples Appendix B WindowingIntroduction Appendix B Windowing B-4 Page