Anritsu S113C, S332C manual Setting the Scale, Save and Recall a Setup, Auto Scale, Amplitude Scale

Page 48

Chapter 3 Getting Started

Setting the Scale

Auto Scale

The Site Master can automatically set the scales to the minimum and maximum values of the measurement on the y-axis of the display. This function is particularly useful for mea- surements in SWR mode. To automatically set the scales:

Step 1. Press the AUTO SCALE key.

The Site Master will automatically set the top and bottom scales to the minimum and maximum values of the measurement on the y-axis of the display.

Amplitude Scale

The following procedure sets the top and bottom scale display.

Step 2. Press the AMPLITUDE key to call up the Scale Menu.

Step 3. Press the TOP soft key and use the keypad or Up/Down arrow key to edit the top scale value.

Step 4. Press the BOTTOM soft key and use the keypad or Up/Down arrow key to edit the bottom scale value.

NOTE: Typically the y-axis scale of the display is 0-54 dB (return loss) but for some measurements (for example, insertion loss) the scale should be changed to 0-10 dB. If the scale is not changed, some measurement results may not be easily displayed on the screen.

Save and Recall a Setup

Saving a Setup

Saving a setup configuration in memory will preserve the calibration information.

Step 1. To save the configuration in one of the 10 available user setup locations, press

SAVE SETUP .

Step 2. Use the key pad or the Up/Down arrow key to select a location (1 - 10). Step 3. Press ENTER to save the setup.

NOTE: A manual calibration setup will be saved with an OSL designator. An InstaCal setup will be saved with an ICAL designation.

3-7

Image 48 Contents
Users Guide Warranty Page Table of Contents Power Measurement Iii/iv Page Standard Accessories Chapter General InformationIntroduction DescriptionOptional Accessories OptionsPerformance Specifications Description Site Master Spectrum Analyzer Preventive Maintenance CalibrationAnnual Verification InstaCal ModuleAnritsu Service Centers Test Connector Panel Chapter Functions and OperationsFront Panel Overview Active Function Block Soft KeysFunction Hard Keys ModeKeypad Hard Keys ClearAuto Soft Keys MODE=FREQ FREQ/DIST SoftkeysMODE=DTF FREQ/DIST Softkeys MODE=SPECTRUM Analyzer FREQ/DIST Amplitude Softkeys Power Monitor Mode Soft Key Labels Functions and Operations Functions and Operations Amplitude RUN/HOLD button Functions and Operations Power Monitor Menu Functions and Operations Limit SYS Symbols Error CodesSelf Test Errors Self TestError Message Range Errors ERRORD2InstaCal Error Messages InstalledError MessageDescription Battery Information Charging a New BatteryCharging the Battery in the Site Master Charging the Battery in the Optional ChargerBattery Charge Indicator Battery LifeImportant Battery Information Power On Procedure Chapter Getting StartedSelect the Frequency or Distance Calibration Trace Characteristics in Return Loss Mode Calibration VerificationManual Calibration Procedure CalibrationInstaCal Module Verification Termination MethodOffset Method InstaCal Module Calibration Procedure Calibration with the Test Port Extension CableSetting the Scale Save and Recall a SetupSaving a Setup Auto ScaleRecalling a Setup Save and Recall a DisplaySaving a Display Recalling a DisplaySet the Distance and Cable Type Changing the UnitsChanging the Display Language Printing Printing a ScreenPrinter Switch Settings PrinterUsing the Soft Carrying Case Using the Site Master Soft Carrying CaseChapter Cable & Antenna Measurements Line Sweep FundamentalsInformation Required for a Line Sweep Cable & Antenna MeasurementsTypical Line Sweep Test Procedures System Return Loss MeasurementProcedure Cable Loss Mode Insertion Loss MeasurementSet M1 to Marker to Peak Set M2 to Marker to Valley Procedure Return Loss Mode Typical Transmission Line Cable Loss TraceTypical Return Loss Trace Distance-To-Fault DTF Transmission Line Test Procedure DTF-SWR Mode Shows an example of a typical DTF return loss traceAntenna Subsystem Return Loss Test 11/4-12 Page Chapter Spectrum Analyzer Measurements Occupied BandwidthChannel Power Measurement Channel Power Measurement with the Site MasterProcedure Adjacent Channel Power Measurement Example Acpr SpecificationCW Wave Out-of-Band Spurious Emission Measurements In-Band/Out-of-Channel Measurements Field Strength Measurement Antenna Calculations VdBV PdBmPage Chapter Power Measurement Power MeasurementDisplaying Relative Power Features Chapter Site Master Software ToolsSystem Requirements Installation Communication Port SettingUsing Software Tools Interface Cable InstallationDownloading Traces from the Site Master Display Mode Plot CapturePlot Properties Graph TitlesTrace Overlay or Plot Overlay Saving TracesMarkers MiscCustom Cable List Saving a Plot to a SpreadsheetCreating a Custom Cable List Entering Antenna Factors Uploading a Cable ListQuerying the Site Master Cable List Uploading Antenna Factors Creating a DatabasePrinting Formats Coaxial Cable Technical Data Appendix a Reference DataAppendix a Reference Data A-4 Page Appendix B Windowing IntroductionExamples Appendix B Windowing B-4 Page