Anritsu S331C, S113C, S332C Procedure Return Loss Mode, Typical Transmission Line Cable Loss Trace

Page 59

Chapter 4 Cable & Antenna Measurements

Figure 4-3 is an example of a typical transmission line cable loss trace.

Figure 4-3. Typical Transmission Line Cable Loss Trace

Procedure - Return Loss Mode

Step 1. Press the MODE key.

Step 2. Select FREQ-RETURN LOSS using the Up/Down arrow key and press

ENTER.

Step 3. Set the start and stop frequencies, F1 and F2, as described on page 3-2.

Step 4. Connect the Test Port Extension cable to the RF port and calibrate the Site Mas- ter as described on page 3-2.

Step 5. Save the calibration set up (page 3-7).

Step 6. Connect the Device Under Test to the Site Master phase stable Test Port Exten- sion cable. A trace will be displayed on the screen as long as the Site Master is in sweep mode.

Step 7. Press the AMPLITUDE key and set TOP and BOTTOM values of the display. In the example below, the TOP is set to 4, and the BOTTOM is set to 10.

Step 8. Press the MARKER key.

Step 9. Set M1 to MARKER TO PEAK.

Step 10. Set M2 to MARKER TO VALLEY.

Step 11. Calculate the insertion loss by averaging M1 (marker to peak) and M2 (marker to valley) and dividing by two as follows:

M1￿ M2

Insertion Loss￿ 2

2

Step 12. Press SAVE DISPLAY (page 3-8) name the trace, and press ENTER.

4-6

Image 59 Contents
Users Guide Warranty Page Table of Contents Power Measurement Iii/iv Page Description Standard AccessoriesChapter General Information IntroductionOptions Optional AccessoriesPerformance Specifications Description Site Master Spectrum Analyzer Calibration Preventive MaintenanceInstaCal Module Annual VerificationAnritsu Service Centers Chapter Functions and Operations Test Connector PanelActive Function Block Soft Keys Front Panel OverviewMode Function Hard KeysClear Keypad Hard KeysAuto MODE=FREQ FREQ/DIST Softkeys Soft KeysMODE=DTF FREQ/DIST Softkeys MODE=SPECTRUM Analyzer FREQ/DIST Amplitude Softkeys Power Monitor Mode Soft Key Labels Functions and Operations Functions and Operations Amplitude RUN/HOLD button Functions and Operations Power Monitor Menu Functions and Operations Limit SYS Self Test SymbolsError Codes Self Test ErrorsError Message ERRORD2 Range ErrorsInstalled InstaCal Error MessagesError MessageDescription Charging the Battery in the Optional Charger Battery InformationCharging a New Battery Charging the Battery in the Site MasterBattery Life Battery Charge IndicatorImportant Battery Information Chapter Getting Started Power On ProcedureCalibration Select the Frequency or DistanceCalibration Verification Trace Characteristics in Return Loss ModeCalibration Manual Calibration ProcedureOffset Method InstaCal Module VerificationTermination Method Calibration with the Test Port Extension Cable InstaCal Module Calibration ProcedureAuto Scale Setting the ScaleSave and Recall a Setup Saving a SetupRecalling a Display Recalling a SetupSave and Recall a Display Saving a DisplayChanging the Display Language Set the Distance and Cable TypeChanging the Units Printing a Screen PrintingPrinter Printer Switch SettingsUsing the Site Master Soft Carrying Case Using the Soft Carrying CaseLine Sweep Fundamentals Chapter Cable & Antenna Measurements Cable & Antenna Measurements Information Required for a Line SweepSystem Return Loss Measurement Typical Line Sweep Test ProceduresInsertion Loss Measurement Procedure Cable Loss ModeSet M1 to Marker to Peak Set M2 to Marker to Valley Typical Transmission Line Cable Loss Trace Procedure Return Loss ModeTypical Return Loss Trace Distance-To-Fault DTF Transmission Line Test Shows an example of a typical DTF return loss trace Procedure DTF-SWR ModeAntenna Subsystem Return Loss Test 11/4-12 Page Occupied Bandwidth Chapter Spectrum Analyzer MeasurementsChannel Power Measurement with the Site Master Channel Power MeasurementProcedure Example Acpr Specification Adjacent Channel Power MeasurementCW Wave Out-of-Band Spurious Emission Measurements In-Band/Out-of-Channel Measurements Field Strength Measurement VdBV PdBm Antenna CalculationsPage Power Measurement Chapter Power MeasurementDisplaying Relative Power System Requirements FeaturesChapter Site Master Software Tools Communication Port Setting InstallationDownloading Traces from the Site Master Using Software ToolsInterface Cable Installation Graph Titles Display ModePlot Capture Plot PropertiesMisc Trace Overlay or Plot OverlaySaving Traces MarkersCreating a Custom Cable List Custom Cable ListSaving a Plot to a Spreadsheet Querying the Site Master Cable List Entering Antenna FactorsUploading a Cable List Printing Formats Uploading Antenna FactorsCreating a Database Appendix a Reference Data Coaxial Cable Technical DataAppendix a Reference Data A-4 Page Examples Appendix B WindowingIntroduction Appendix B Windowing B-4 Page