Chapter 3 Getting Started

Set the Distance and Cable Type

In Distance-To-Fault (DTF) mode, the length of the transmission line (distance) and cable type are selected. The cable type determines the velocity propagation and cable attenuation factor. The following procedure can be used to set the distance and select the appropriate cable type.

NOTE: Selecting the correct cable is very important for accurate measurements and for identifying faults in the transmission line. Selecting the incorrect cable type will shift the DTF trace vertically and horizontally making it difficult to accu- rately locate faults.

Step 1. Press the MODE key.

Step 2. Select DTF RETURN LOSS or DTF SWR mode. The Site Master automatically sets D1 to zero.

Step 3. Press the D2 soft key.

Step 4. Enter the appropriate D2 value for the maximum length of the transmission line and press the ENTER key to set the D2 value.

Step 5. Press the DTF AID soft key.

Step 6. Using the Up/Down arrow key, select CABLE TYPE. Standard cable types are stored in the Site Master cable list. This standard list may not be edited. A cus- tom cable list can also be created.

Step 7. Using the Up/Down arrow key, select the appropriate list and select the type of cable. The selected cable type, PROP VEL and CABLE LOSS in dB/ft (or dB/m) will be displayed.

Step 8. Press ENTER.

Changing the Units

By default, the Site Master displays information in metric units. Use the following proce- dure to change the display to English units.

Step 1. Press the SYS key.

Step 2. Select the OPTIONS soft key.

Step 3. Press UNITS to change from metric to English measurement units, or vice versa. The current selection is displayed at the bottom left corner of the screen.

Changing the Display Language

By default, the Site Master displays messages in English. To change the display language:

Step 1. Press the SYS key.

Step 2. Select the Language soft key.

Step 3. Select the desired language. Choices are English, French, German, Spanish, Chi- nese, and Japanese. The default language is English.

3-9

Image 50 Contents
Users Guide Warranty Page Table of Contents Power Measurement Iii/iv Page Introduction Standard AccessoriesChapter General Information DescriptionOptional Accessories OptionsPerformance Specifications Description Site Master Spectrum Analyzer Preventive Maintenance CalibrationAnnual Verification InstaCal ModuleAnritsu Service Centers Test Connector Panel Chapter Functions and OperationsFront Panel Overview Active Function Block Soft KeysFunction Hard Keys ModeKeypad Hard Keys ClearAuto Soft Keys MODE=FREQ FREQ/DIST SoftkeysMODE=DTF FREQ/DIST Softkeys MODE=SPECTRUM Analyzer FREQ/DIST Amplitude Softkeys Power Monitor Mode Soft Key Labels Functions and Operations Functions and Operations Amplitude RUN/HOLD button Functions and Operations Power Monitor Menu Functions and Operations Limit SYS Self Test Errors SymbolsError Codes Self TestError Message Range Errors ERRORD2InstaCal Error Messages InstalledError MessageDescription Charging the Battery in the Site Master Battery InformationCharging a New Battery Charging the Battery in the Optional ChargerBattery Charge Indicator Battery LifeImportant Battery Information Power On Procedure Chapter Getting StartedSelect the Frequency or Distance CalibrationTrace Characteristics in Return Loss Mode Calibration VerificationManual Calibration Procedure Calibration Offset Method InstaCal Module Verification Termination Method InstaCal Module Calibration Procedure Calibration with the Test Port Extension CableSaving a Setup Setting the ScaleSave and Recall a Setup Auto ScaleSaving a Display Recalling a SetupSave and Recall a Display Recalling a DisplayChanging the Display Language Set the Distance and Cable TypeChanging the Units Printing Printing a ScreenPrinter Switch Settings PrinterUsing the Soft Carrying Case Using the Site Master Soft Carrying CaseChapter Cable & Antenna Measurements Line Sweep FundamentalsInformation Required for a Line Sweep Cable & Antenna MeasurementsTypical Line Sweep Test Procedures System Return Loss MeasurementProcedure Cable Loss Mode Insertion Loss MeasurementSet M1 to Marker to Peak Set M2 to Marker to Valley Procedure Return Loss Mode Typical Transmission Line Cable Loss TraceTypical Return Loss Trace Distance-To-Fault DTF Transmission Line Test Procedure DTF-SWR Mode Shows an example of a typical DTF return loss traceAntenna Subsystem Return Loss Test 11/4-12 Page Chapter Spectrum Analyzer Measurements Occupied BandwidthChannel Power Measurement Channel Power Measurement with the Site MasterProcedure Adjacent Channel Power Measurement Example Acpr SpecificationCW Wave Out-of-Band Spurious Emission Measurements In-Band/Out-of-Channel Measurements Field Strength Measurement Antenna Calculations VdBV PdBmPage Chapter Power Measurement Power MeasurementDisplaying Relative Power System Requirements FeaturesChapter Site Master Software Tools Installation Communication Port SettingDownloading Traces from the Site Master Using Software ToolsInterface Cable Installation Plot Properties Display ModePlot Capture Graph TitlesMarkers Trace Overlay or Plot OverlaySaving Traces MiscCreating a Custom Cable List Custom Cable ListSaving a Plot to a Spreadsheet Querying the Site Master Cable List Entering Antenna FactorsUploading a Cable List Printing Formats Uploading Antenna FactorsCreating a Database Coaxial Cable Technical Data Appendix a Reference DataAppendix a Reference Data A-4 Page Examples Appendix B WindowingIntroduction Appendix B Windowing B-4 Page