Page 24

Data management

There are two data ports at the rear of the device for connecting a USB stick (see “Stellar at a glance” on page 2). Therapy and device data can be stored on it or read from it to be used with ResMed software applications.

WARNING

Do not connect any device to the data ports other than specially designed devices recommended by ResMed. Connection of other devices could result in injury, or damage to the Stellar (see “General warnings and cautions” on page 30).

CAUTION

Do not disconnect the ResMed USB stick while data transfer is in progress. Otherwise loss of data or incorrect data may result. The download time depends on the data volume.

Notes:

You cannot connect two ResMed USB sticks simultaneously for data communication.

If data transfer is not possible or failed please read the troubleshooting section.

Do not store files on the USB stick other than those created by the device or application. Unknown files may be lost during data transfer.

1 Plug the ResMed USB stick to one of the two USB connections on the rear side of the device.

The first dialogue for the USB data transfer is displayed automatically on the LCD. The device checks if there is enough memory capacity on the USB stick and if readable data is available.

2Select one of the available options

Read settings

The settings will be transfered from the connected ResMed USB stick to the Stellar.

Write settings

Settings of the device will be stored on the ResMed USB stick.

Write settings and loggings

Settings and loggings of the device will be stored on the ResMed USB stick.

Cancel

3Confirm the data transfer.

Cleaning and maintenance

The cleaning and maintenance described in this section should be carried out regularly. This also helps to prevent the risk of cross contamination. Refer to the mask, humidifier and other accessories user guide for detailed instructions for care and maintenance.

WARNING

• Beware of electric shock. Do not immerse the device, pulse oximeter or power cord in water. Turn off the device, unplug the power cord from the power socket and the device before cleaning and be sure that it is dry before reconnecting.

The mask system and air tubing are subject to normal wear and tear. Inspect them regularly for damage.

CAUTION

The device cannot be sterilised.

Daily

Disconnect the air tubing from the device (and humidifier, if used) and hang it in a clean, dry place until next use.

CAUTION

Do not hang the air tubing in direct sunlight as the tubing may harden over time and eventually crack.

English

Data management

19

Image 24
Contents Stellar User GuideRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Cleaning and maintenance Introduction Indications for useContraindications Adverse effectsStellar at a glance Internal battery Patient interfaceHumidification ResMed USB stickSetting up for noninvasive use Mobile useSetting up for invasive use Attaching the H4i heated humidifier for noninvasive useHmef Working with other optional accessories Attaching the pulse oximeterAdding supplemental oxygen Starting therapy using oxygen Stopping therapy using oxygenUsing the FiO2 monitoring sensor Connecting a sensor Attaching an antibacterial filterStellar basics About the control panelStarting therapy LCD screenPerforming a functional test Starting therapy Stopping therapyCheck the FiO2 monitoring sensor if in use Check pulse oximeter if in useTurning off the power Working with alarmsTailoring treatment setup options Using mask-fitUsing the menus Enu Alarm Settings Alarm Volume Enu OptionsSetup menu Configuration Menu Parameter DescriptionOptions On, Auto Info menu Event SummaryUsed Hours Device InformationCleaning and maintenance Data managementDaily Weekly MonthlyReplacing the air filter DisinfectionTroubleshooting Alarm troubleshootingMultipatient use ServicingProblem / possible cause Action LCD Internal Battery Empty LCD High Temperature 10, 11, 12LCD System Failure LCD Over PressureProblem / possible cause Action LCD Circuit disconnected LCD Internal Battery LowLCD Low Minute Ventilation LCD Low Respiratory Rate!, High Respiratory RateOther troubleshooting Problem / possible cause Solution Device does not start when you breathe into the maskDevice does not stop when you remove your mask Treatment pressure seems highTechnical specifications Alarm Volume Range Dimensions L x W x HWeight Air outlet Power supplyIEC ClassificationsAir travel requirements Vrms Symbols General warnings and cautionsGeneral warnings and cautions Limited warranty Product Warranty periodApac