ResMed 2011-09 Problem / possible cause Action LCD Internal Battery Empty, LCD System Failure

Page 27

Problem / possible cause

Action

 

 

LCD: Internal Battery Empty!

The remaining battery charge is below 15%.The device can be powered by the internal battery for maximum 2 minutes.

Connect the device to mains power. Note: In case of a total power failure, the therapy settings will be stored and therapy will resume when the device is powered again.

LCD: System Failure!

Component failure.

The device stops delivering air pressure (system failure 6, 7, 9, 22, 38). Therapy cannot be started (system failure 21).

1.Power-off the device.

2.Power-on the device again.

Component failure (system failure 8, 25).

1.

Power-off the device.

 

2.

Power-on the device again.

LCD: Over Pressure!

The device generates a pressure that is greater than 60 cm H2O. Treatment will be stopped.

1.Power-off the device.

2.Check that the air tubing is connected properly.

3.Power-on the device again.

4.Start Learn Circuit function.

Note: If the alarm activates repeatedly internal components may be defective. Discontinue use and return the device for servicing.

LCD: Blocked Tube!

Air path is blocked.

1.

Check the air path for any blockages.

 

2.

Remove the blockages.

 

3.

If the alarm is not cleared, stop

 

 

treatment.

 

4.

Re-start treatment.

LCD: High Temperature [10, 11, 12, 13]!

The temperature inside the device is too high. Treatment may lead to stop.

Ensure the ambient temperature is within the specified operating range. If the problem persists within the specified operating conditions please return the device for servicing.

Contact your clinician.

LCD: High Pressure!

Therapy pressure exceeds pre-set alarm level.

1.Stop treatment.

2.Re-start treatment.

If the problem persists contact your clinician.

LCD: Low Pressure!

The air tubing is not connected properly.

1.Check the air circuit integrity and reconnect.

2.If the alarm doesn‘t get cleared, stop treatment.

3.Re-start treatment.

22

Image 27
Contents User Guide StellarRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Cleaning and maintenance Adverse effects IntroductionIndications for use ContraindicationsStellar at a glance ResMed USB stick Internal batteryPatient interface HumidificationMobile use Setting up for noninvasive useAttaching the H4i heated humidifier for noninvasive use Setting up for invasive useHmef Working with other optional accessories Attaching the pulse oximeterAdding supplemental oxygen Starting therapy using oxygen Stopping therapy using oxygenUsing the FiO2 monitoring sensor Attaching an antibacterial filter Connecting a sensorAbout the control panel Stellar basicsStarting therapy LCD screenPerforming a functional test Check pulse oximeter if in use Starting therapyStopping therapy Check the FiO2 monitoring sensor if in useWorking with alarms Turning off the powerUsing mask-fit Tailoring treatment setup optionsUsing the menus Enu Options Enu Alarm Settings Alarm VolumeSetup menu Configuration Menu Parameter DescriptionOptions On, Auto Device Information Info menuEvent Summary Used HoursCleaning and maintenance Data managementDaily Disinfection WeeklyMonthly Replacing the air filterServicing TroubleshootingAlarm troubleshooting Multipatient useLCD Over Pressure Problem / possible cause Action LCD Internal Battery EmptyLCD High Temperature 10, 11, 12 LCD System FailureLCD Low Respiratory Rate!, High Respiratory Rate Problem / possible cause Action LCD Circuit disconnectedLCD Internal Battery Low LCD Low Minute VentilationOther troubleshooting Treatment pressure seems high Problem / possible cause SolutionDevice does not start when you breathe into the mask Device does not stop when you remove your maskTechnical specifications Power supply Alarm Volume RangeDimensions L x W x H Weight Air outletIEC ClassificationsAir travel requirements Vrms General warnings and cautions SymbolsGeneral warnings and cautions Product Warranty period Limited warrantyApac