Multipatient use

WARNING

An antibacterial filter is mandatory if the device is used on multiple patients.

In a mulitpatient use environment, you must perform the following before the device is provided to a new patient:

Air filter and the

Replace.

antibacterial filter

 

 

Mask

Reprocess; Cleaning, disinfection and sterilisation

 

instructions are available from the ResMed website,

 

www.resmed.com/masks/sterilization. If you do not have

 

Internet access, please contact your ResMed representative.

Air tubing

Replace the air tubing. Alternatively, consult the air tubing

 

instructions for cleaning and disinfection information.

Device

Disinfect the Stellar as follows:

 

1.

Apply undiluted mikrozid® AF or CaviCide® liquid to a

 

 

clean non-dyed disposable cloth.

 

2.

Wipe all surfaces of the device, including the air outlet;

 

 

avoid liquid entering any openings in the device.

 

3.

Leave the disinfectant to work for five minutes.

 

4.

Wipe residual disinfectant from the device with a clean,

 

 

dry non-dyed disposable cloth.

Humidifier

As instructions for humidifiers vary, see the user guide for

 

the humidifier in use.

Servicing

CAUTION

Inspection and repair should only be performed by an authorised agent. Under no circumstances should you attempt to open, service or repair the device yourself.

This product should be inspected by an authorised ResMed service centre five years from the date of manufacture, except for the internal battery which must be replaced every two years from the manufacturing date of the device. Prior to this, the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed. Applicable ResMed warranty details are provided with the device at the time of original supply. Of course, as with all electrical devices, if any irregularity becomes apparent, you should exercise caution and have the device inspected by an authorised ResMed service centre.

Troubleshooting

If there is a problem, try the following suggestions. If a problem cannot be solved, contact ResMed.

Alarm troubleshooting

The most common reason for an alarm to sound is because the system has not been properly assembled. Check that the air tubing has been properly attached to the device and patient interface (and humidifier if used).

Notes:

The alarm log and alarm settings are maintained when the device is powered down and in the event of a power loss.

If multiple alarms are active simultaneously, the alarm with the highest priority will be displayed first.

If an alarm activates repeatedly, discontinue use and return the device for servicing.

English

Troubleshooting 21

Image 26
Contents Stellar User GuideRespiratory Care Solutions Making quality of care easy Stellar 100 Stellar Contents Cleaning and maintenance Contraindications IntroductionIndications for use Adverse effectsStellar at a glance Humidification Internal batteryPatient interface ResMed USB stickSetting up for noninvasive use Mobile useSetting up for invasive use Attaching the H4i heated humidifier for noninvasive useHmef Adding supplemental oxygen Working with other optional accessoriesAttaching the pulse oximeter Using the FiO2 monitoring sensor Starting therapy using oxygenStopping therapy using oxygen Connecting a sensor Attaching an antibacterial filterStellar basics About the control panelPerforming a functional test Starting therapyLCD screen Check the FiO2 monitoring sensor if in use Starting therapyStopping therapy Check pulse oximeter if in useTurning off the power Working with alarmsTailoring treatment setup options Using mask-fitUsing the menus Enu Alarm Settings Alarm Volume Enu OptionsOptions On, Auto Setup menu Configuration MenuParameter Description Used Hours Info menuEvent Summary Device InformationDaily Cleaning and maintenanceData management Replacing the air filter WeeklyMonthly DisinfectionMultipatient use TroubleshootingAlarm troubleshooting ServicingLCD System Failure Problem / possible cause Action LCD Internal Battery EmptyLCD High Temperature 10, 11, 12 LCD Over PressureLCD Low Minute Ventilation Problem / possible cause Action LCD Circuit disconnectedLCD Internal Battery Low LCD Low Respiratory Rate!, High Respiratory RateOther troubleshooting Device does not stop when you remove your mask Problem / possible cause SolutionDevice does not start when you breathe into the mask Treatment pressure seems highTechnical specifications Weight Air outlet Alarm Volume RangeDimensions L x W x H Power supplyAir travel requirements IECClassifications Vrms Symbols General warnings and cautionsGeneral warnings and cautions Limited warranty Product Warranty periodApac