ResMed Stellar 100 and 150 manual Starting therapy, LCD screen, Performing a functional test

Page 15

LCD screen

The LCD screen displays the menus, monitoring screens and alarm status.

Values on this LCD screen and other screens throughout the User Guide are examples only.

Warm-up feature H4i

ResMed USB stick Mains power

External power supply

Internal battery

Title bar

Settings

Therapy status bar

Ramp time

Program

Therapy mode

Patient mode or Clinical mode

Current screen/number of screens within the menu

Scroll bar

Starting therapy

Performing a functional test

Make sure your device is functioning properly each time before starting therapy.

If any problems occur, see “Troubleshooting” on page 22. Please also check other provided User Instructions for troubleshooting information.

1Turn off the device by pressing the power switch at the back of the device.

2Check condition of device and accessories.

Inspect the device and all the provided accessories. If there are any visible defects, the system should not be used.

3Check the circuit configuration.

Check the integrity of the circuit configuration (device and provided accessories) according to the setup descriptions in this User Guide and that all connections are secure.

4Turn on the device and check alarms.

Press the power switch at the back of the device once to turn on the device.

Check that the alarm sounds a test beep and the LEDs (visual indicator) for the alarm signal and the Alarm mute button flash. The device is ready for use when the Treatment screen

is displayed. If the display shows the Reminder page, follow the instructions, then press

to display the Treatment screen.

5Check batteries.

Disconnect the device from the mains and external battery (if in use) so that the device is powered by the internal battery. Check that the Battery use alarm is displayed and the battery LED is on.

Note: If the charge state of the internal battery is too low, or if the battery is empty, an alarm occurs. See the Alarm troubleshooting section on page 22 for further information.

Reconnect the external battery (if in use) and check that the LED for the external power supply is lit. The External DC power use alarm will be displayed and the Alarm LED will light.

Reconnect the device to the mains.

Starting therapy

11

Image 15
Contents User Guide StellarRespiratory Care Solutions Making quality of care easy Contents Data management Cleaning and maintenance Adverse effects IntroductionIndications for use ContraindicationsStellar at a glance ResMed USB stick Internal batteryPatient interface HumidificationSetting up for noninvasive use Remote AlarmMobile use Attaching the H4i heated humidifier for noninvasive use Setting up for invasive useHmef Working with other optional accessories Attaching the pulse oximeterAdding supplemental oxygen Starting therapy using oxygen Stopping therapy using oxygen Using the FiO2 monitoring sensor Attaching an antibacterial filter Connecting a sensorAbout the control panel Stellar basicsStarting therapy LCD screenPerforming a functional test Stopping therapy Starting therapyWorking with alarms Turning off the powerUsing mask-fit Tailoring treatment setup optionsUsing the menus Setup menu Options Setup menuSetup menu Clinical Settings Mask Type Setup menu Alarm Settings Alarm VolumeSetup menu Configuration Menu Parameter DescriptionOptions On, Auto Device Information Info menuEvent Summary Used HoursCleaning and maintenance Data managementDaily Weekly MonthlyReplacing the air filter Multipatient use DisinfectionTroubleshooting Alarm troubleshootingServicing LCD Low Pressure Problem / possible cause Action LCD Blocked TubeLCD High Temperature 10, 11, 12 LCD High PressureLCD Apnoea Problem / possible cause Action LCD Non-Vented MaskLCD Internal Battery Low LCD Internal Battery UseOther troubleshooting USB stick is not readable or writeable Problem / possible cause SolutionDevice does not stop when you remove your mask Learn Circuit failedTechnical specifications Housing construction Power supplyExternal DC Power Supply isolated Internal BatteryEnglish Vrms Symbols Mask should never be blocked General warnings and cautionsGeneral warnings and cautions Product Warranty period Limited warrantyRespiratory Care Solutions Making quality of care easy EUR1