ResMed Stellar 100 and 150 Setup menu Configuration Menu, Parameter Description, Options On, Auto

Page 21

Parameter

Description

 

 

Learn Circuit

The device is calibrated according to your air tubing system.

 

1.

If in use, turn off the oxygen flow.

 

2.

Select the mask type.

 

3.

Set up the air circuit including accessories and patient interface.

 

 

Note: For invasive use, do not connect a catheter mount,

 

 

tracheostomy tube or HMEF, see “Setting up for invasive use”

 

 

on page 5).

 

4.

Leave the air circuit unobstructed and open to the air.

 

5.

Press

to start the Learn Circuit.

 

6.

Wait for the device to complete its automated tests (<30 sec).

 

 

The results are displayed when complete. If the circuit

 

 

configuration has been successfully learnt, displays. If

 

 

unsuccessful, displays (see “Troubleshooting” on page 22).

 

 

FiO2 Sensor

The device starts the calibration of the FiO2 monitoring sensor to

Calibration

measure the oxygen concentration of the breathable air.

 

1.

Press

to start the FiO2 sensor calibration.

 

2.

Wait for the device to complete its calibration.

The results are displayed when complete.

Note: Turn off the oxygen flow.

Ramp Time

If Max Ramp Time has been set by the clinician, you may select any

 

value up to this time.

 

Options: 0 min-Max Ramp Time (max 45 minutes, 5 minutes

 

increments)

 

 

Setup menu: Configuration Menu

Parameter

Description

 

 

Language

Sets the display language.

 

Options: Depending on regional configuration

 

 

Brightness

Sets the LCD backlight brightness.

 

Options: 20-100%, 10% increments

 

 

Backlight

Enables the LCD and keypad backlight.

 

If the AUTO setting is selected the backlight turns dark after five

 

minutes without any actions and turns on again if any button is

 

pressed again or an alarm occurs.

 

Options: On, Auto

 

 

Time format

Sets the time format.

 

Options: 24 hrs,12 hrs

 

 

Date format

Sets the date format.

 

Options: dd/mm/yyyy, mm/dd/yyyy

To return to the Options screen press the Return button

English

Setup menu

17

Image 21
Contents User Guide StellarRespiratory Care Solutions Making quality of care easy Contents Data management Cleaning and maintenance Indications for use IntroductionContraindications Adverse effectsStellar at a glance Patient interface Internal batteryHumidification ResMed USB stickSetting up for noninvasive use Remote AlarmMobile use Attaching the H4i heated humidifier for noninvasive use Setting up for invasive useHmef Working with other optional accessories Attaching the pulse oximeterAdding supplemental oxygen Starting therapy using oxygen Stopping therapy using oxygenUsing the FiO2 monitoring sensor Attaching an antibacterial filter Connecting a sensorAbout the control panel Stellar basicsStarting therapy LCD screenPerforming a functional test Stopping therapy Starting therapyWorking with alarms Turning off the power Using mask-fit Tailoring treatment setup optionsUsing the menus Setup menu Clinical Settings Mask Type Setup menuSetup menu Alarm Settings Alarm Volume Setup menu OptionsSetup menu Configuration Menu Parameter DescriptionOptions On, Auto Event Summary Info menuUsed Hours Device InformationCleaning and maintenance Data managementDaily Weekly MonthlyReplacing the air filter Multipatient use DisinfectionTroubleshooting Alarm troubleshootingServicing LCD High Temperature 10, 11, 12 Problem / possible cause Action LCD Blocked TubeLCD High Pressure LCD Low PressureLCD Internal Battery Low Problem / possible cause Action LCD Non-Vented MaskLCD Internal Battery Use LCD ApnoeaOther troubleshooting Device does not stop when you remove your mask Problem / possible cause SolutionLearn Circuit failed USB stick is not readable or writeableTechnical specifications External DC Power Power supplySupply isolated Internal Battery Housing constructionEnglish Vrms Symbols Mask should never be blocked General warnings and cautionsGeneral warnings and cautions Product Warranty period Limited warrantyRespiratory Care Solutions Making quality of care easy EUR1