Air filter
Air filter cover
English |
Disinfection
Disinfection of your device helps to prevent the risk of cross contamination.
Disinfect the exterior of the device, and especially the air outlet, with a damp cloth and a disinfection solution (eg, Microzid).
Multipatient use
WARNING
• An antibacterial filter is mandatory if the device is used on multiple patients.
•In a mulitpatient use environment, you must perform the following before the device is provided to a new patient:
Air filter and the | Replace. |
antibacterial filter |
|
Mask | Reprocess; Cleaning, disinfection and sterilisation |
| instructions are available from the ResMed website, |
| www.resmed.com/masks/sterilization. If you do not have |
| Internet access, please contact your ResMed representative. |
Air tubing | Replace the air tubing. Alternatively, consult the air tubing |
| instructions for cleaning and disinfection information. |
Device | Disinfect the Stellar as follows: |
| Use an |
| Mikrozid® AF or CaviCide® with a clean |
| disposable cloth to clean and disinfect the exterior surfaces |
| of the device. Wipe all accessible surfaces of the device, |
| including the air outlet (avoid liquid entering any openings in |
| the device). Follow the manufacturer’s recommended |
| cleaning instructions. |
Humidifier | As instructions for humidifiers vary, see the user guide for |
| the humidifier in use. For multipatient use environment, use |
| the H4i water chamber (disposable) in place of the H4i |
| reusable water chamber. |
Cleaning and maintenance | 21 |