ResMed Understanding the Importance of Antibacterial Filter and Safe Device Use

Page 25

.

Air filter

Air filter cover

English

Disinfection

Disinfection of your device helps to prevent the risk of cross contamination.

Disinfect the exterior of the device, and especially the air outlet, with a damp cloth and a disinfection solution (eg, Microzid).

Multipatient use

WARNING

• An antibacterial filter is mandatory if the device is used on multiple patients.

In a mulitpatient use environment, you must perform the following before the device is provided to a new patient:

Air filter and the

Replace.

antibacterial filter

 

Mask

Reprocess; Cleaning, disinfection and sterilisation

 

instructions are available from the ResMed website,

 

www.resmed.com/masks/sterilization. If you do not have

 

Internet access, please contact your ResMed representative.

Air tubing

Replace the air tubing. Alternatively, consult the air tubing

 

instructions for cleaning and disinfection information.

Device

Disinfect the Stellar as follows:

 

Use an anti-bacterial cleaning/disinfectant solution such as

 

Mikrozid® AF or CaviCide® with a clean non-dyed

 

disposable cloth to clean and disinfect the exterior surfaces

 

of the device. Wipe all accessible surfaces of the device,

 

including the air outlet (avoid liquid entering any openings in

 

the device). Follow the manufacturer’s recommended

 

cleaning instructions.

Humidifier

As instructions for humidifiers vary, see the user guide for

 

the humidifier in use. For multipatient use environment, use

 

the H4i water chamber (disposable) in place of the H4i

 

reusable water chamber.

Cleaning and maintenance

21

Image 25
Contents User Guide StellarRespiratory Care Solutions Making quality of care easy Contents Data management Cleaning and maintenance Indications for use IntroductionContraindications Adverse effectsStellar at a glance Patient interface Internal batteryHumidification ResMed USB stickRemote Alarm Setting up for noninvasive useMobile use Attaching the H4i heated humidifier for noninvasive use Setting up for invasive useHmef Attaching the pulse oximeter Working with other optional accessoriesAdding supplemental oxygen Stopping therapy using oxygen Starting therapy using oxygenUsing the FiO2 monitoring sensor Attaching an antibacterial filter Connecting a sensorAbout the control panel Stellar basicsLCD screen Starting therapyPerforming a functional test Stopping therapy Starting therapyWorking with alarms Turning off the powerUsing mask-fit Tailoring treatment setup optionsUsing the menus Setup menu Clinical Settings Mask Type Setup menuSetup menu Alarm Settings Alarm Volume Setup menu OptionsParameter Description Setup menu Configuration MenuOptions On, Auto Event Summary Info menuUsed Hours Device InformationData management Cleaning and maintenanceDaily Monthly WeeklyReplacing the air filter Multipatient use DisinfectionAlarm troubleshooting TroubleshootingServicing LCD High Temperature 10, 11, 12 Problem / possible cause Action LCD Blocked TubeLCD High Pressure LCD Low PressureLCD Internal Battery Low Problem / possible cause Action LCD Non-Vented MaskLCD Internal Battery Use LCD ApnoeaOther troubleshooting Device does not stop when you remove your mask Problem / possible cause SolutionLearn Circuit failed USB stick is not readable or writeableTechnical specifications External DC Power Power supplySupply isolated Internal Battery Housing constructionEnglish Vrms Symbols Mask should never be blocked General warnings and cautionsGeneral warnings and cautions Product Warranty period Limited warrantyRespiratory Care Solutions Making quality of care easy EUR1