Stellar basics

About the control panel

l

Mains Power LED

LCD Screen

On when operating on mains power.

External power supply

LED

On when external battery is connected.

Internal battery LED

On when internal battery is in use. Flashes when device is turned off and battery is charging.

Alarm LED

Monitoring menu

Setup menu

Info menu

Red or yellow during an alarm or during alarm testing.

Alarm mute button LED

On when is pressed.

Start/stop button

Key

Start/Stop

Push dial (rotate/

click)

Therapy LED

On during treatment. Flashes during mask- fitting operation.

Function

Starts or stops treatment.

Extended hold for at least three seconds starts the mask­fitting feature.

Alarm mute

• During therapy: Press once to mute an alarm. Press a

 

 

 

second time to un-mute an alarm. If the problem is still

 

 

 

present, the alarm will sound again after two minutes.

 

 

 

See “Working with alarms” on page 13.

In stand-by mode: Extended hold for at least three seconds starts the alarm LED and alarm buzzer test.

Menu keys

Press the appropriate menu button (Monitoring, Setup, Info)

 

to enter and scroll through the relevant menu.

Push dial

Turning the dial allows you to scroll through the menu and

 

change settings. Pushing the dial allows you to enter into a

 

menu or confirm your choice.

10

Image 14
Contents Stellar User GuideRespiratory Care Solutions Making quality of care easy Contents Data management Cleaning and maintenance Contraindications IntroductionIndications for use Adverse effectsStellar at a glance Humidification Internal batteryPatient interface ResMed USB stickMobile use Setting up for noninvasive useRemote Alarm Setting up for invasive use Attaching the H4i heated humidifier for noninvasive useHmef Adding supplemental oxygen Working with other optional accessories Attaching the pulse oximeter Using the FiO2 monitoring sensor Starting therapy using oxygenStopping therapy using oxygen Connecting a sensor Attaching an antibacterial filterStellar basics About the control panelPerforming a functional test Starting therapyLCD screen Starting therapy Stopping therapyTurning off the power Working with alarmsTailoring treatment setup options Using mask-fitUsing the menus Setup menu Alarm Settings Alarm Volume Setup menuSetup menu Clinical Settings Mask Type Setup menu OptionsOptions On, Auto Setup menu Configuration MenuParameter Description Used Hours Info menuEvent Summary Device InformationDaily Cleaning and maintenanceData management Replacing the air filter WeeklyMonthly Disinfection Multipatient useServicing TroubleshootingAlarm troubleshooting LCD High Pressure Problem / possible cause Action LCD Blocked TubeLCD High Temperature 10, 11, 12 LCD Low PressureLCD Internal Battery Use Problem / possible cause Action LCD Non-Vented MaskLCD Internal Battery Low LCD ApnoeaOther troubleshooting Learn Circuit failed Problem / possible cause SolutionDevice does not stop when you remove your mask USB stick is not readable or writeableTechnical specifications Supply isolated Internal Battery Power supplyExternal DC Power Housing constructionEnglish Vrms Symbols General warnings and cautions Mask should never be blockedGeneral warnings and cautions Limited warranty Product Warranty periodRespiratory Care Solutions Making quality of care easy EUR1