Problem / possible cause | Action |
|
|
LCD: Non-Vented Mask!
•Connection of a
•Mask vents may be blocked.
•ResMed Leak Valve is missing or vent is blocked.
•Ensure the mask has vents.
•Ensure the mask vents are not blocked.
•Ensure the ResMed Leak Valve is installed and that the vent is not blocked.
•Ensure oxygen (if in use) has only been connected at the rear of the device.
•If the problem persists contact your clinician.
LCD: Apnoea!
The device detects an apnoea that has exceeded the
•Breathe normally to disable the alarm.
•If the problem persists contact your clinician.
LCD: Internal Battery Low!
The internal battery capacity is below 30%. | Connect the device to mains power. |
LCD: Low SpO2! |
|
SpO2 has dropped below | • Check the attachment of the sensor. |
level. | • If the problem persists contact your |
| clinician. |
|
|
LCD: SpO2 finger sensor failure! |
|
The finger sensor is not connected properly or delivers faulty values.
Check if the finger sensor is attached properly to the finger and connection to the pulse oximeter.
LCD: Xpod oximeter disconnected!
The pulse oximeter is disconnected.
Check if the pulse oximeter is connected properly to the device.
LCD: Low FiO2 Level!
FiO2 has dropped below the
•Perform FiO2 sensor calibration.
•If the problem persists contact your clinician.
LCD: High FiO2 Level!
FiO2 has exceeded the | • Perform FiO2 sensor calibration. |
level. | • If the problem persists contact your |
| clinician. |
|
|
LCD: Keypad Failure! |
|
|
|
One of the keys was held down for more | Remove any blockages from the keypad. |
than 10 seconds or got stuck. |
|
LCD: Attention High Temp [42, 43, 44, 45]!
The temperature inside the device is high.
Ensure the ambient temperature conditions are within the specified operating range.
LCD: Internal Battery Use!
The device is using the internal battery.
Check if the power cord is properly connected to the device if you want to run from mains power.
Press the Alarm mute button to cancel the alarm.