ResMed Stellar 150, Stellar 100 Setup menu Clinical Settings Mask Type, Setup menu Options

Page 20

Setup menu

Setup menu: Clinical Settings (Mask Type)

1Press to display the Clinical Settings screen.

2Use the push dial to scroll through the menu and change the Mask Type in the Advanced Settings screen.

Mask type options: Nasal, Ultra, Pillow, Full Face, Trach, Pediatric.

Note: When Mask Type is set to Trach or Full Face the Non-Vented Mask alarm automatically turns on.

For a full list of compatible masks for this device, see the Mask/Device Compatibility List on www.resmed.com on the Products page under Service & Support. If you do not have internet access, please contact your ResMed representative.

Setup menu: Alarm Settings (Alarm Volume)

1Press to display the Alarm Settings screen.

2Use the push dial to scroll through the menu and change the Alarm Volume to low, medium or high.

Setup menu: Options

1Press to display the Options screen.

2Use the push dial to scroll through the menu and change parameters (see descriptions in the table below).

16

Image 20
Contents Stellar User GuideRespiratory Care Solutions Making quality of care easy Contents Data management Cleaning and maintenance Introduction Indications for useContraindications Adverse effectsStellar at a glance Internal battery Patient interfaceHumidification ResMed USB stickMobile use Setting up for noninvasive useRemote Alarm Setting up for invasive use Attaching the H4i heated humidifier for noninvasive useHmef Adding supplemental oxygen Working with other optional accessoriesAttaching the pulse oximeter Using the FiO2 monitoring sensor Starting therapy using oxygenStopping therapy using oxygen Connecting a sensor Attaching an antibacterial filterStellar basics About the control panelPerforming a functional test Starting therapyLCD screen Starting therapy Stopping therapyTurning off the power Working with alarmsTailoring treatment setup options Using mask-fitUsing the menus Setup menu Setup menu Clinical Settings Mask TypeSetup menu Alarm Settings Alarm Volume Setup menu OptionsOptions On, Auto Setup menu Configuration MenuParameter Description Info menu Event SummaryUsed Hours Device InformationDaily Cleaning and maintenanceData management Replacing the air filter WeeklyMonthly Disinfection Multipatient useServicing TroubleshootingAlarm troubleshooting Problem / possible cause Action LCD Blocked Tube LCD High Temperature 10, 11, 12LCD High Pressure LCD Low PressureProblem / possible cause Action LCD Non-Vented Mask LCD Internal Battery LowLCD Internal Battery Use LCD ApnoeaOther troubleshooting Problem / possible cause Solution Device does not stop when you remove your maskLearn Circuit failed USB stick is not readable or writeableTechnical specifications Power supply External DC PowerSupply isolated Internal Battery Housing constructionEnglish Vrms Symbols General warnings and cautions Mask should never be blockedGeneral warnings and cautions Limited warranty Product Warranty periodRespiratory Care Solutions Making quality of care easy EUR1