ResMed Stellar 100 and 150 Problem / possible cause Action LCD Blocked Tube, LCD High Pressure

Page 27

Problem / possible cause

Action

LCD: Blocked Tube!

 

 

 

 

 

Air path is blocked.

1.

Check the air path for any blockages.

 

2.

Remove the blockages.

 

3.

If the alarm is not cleared, stop

 

 

treatment.

 

4.

Re-start treatment.

 

 

LCD: High Temperature [10, 11, 12, 23]!

 

 

 

The temperature inside the device is too

Ensure the ambient temperature is within

high. Treatment may lead to stop.

the specified operating range. If the

 

problem persists within the specified

 

operating conditions please return the

 

device for servicing.

 

Contact your clinician.

 

 

 

LCD: High Pressure!

 

 

 

 

 

Therapy pressure exceeds pre-set alarm

1.

Stop treatment.

level.

2.

Re-start treatment.

 

If the problem persists contact your

 

clinician.

 

 

 

LCD: Low Pressure!

 

 

 

 

 

The air tubing is not connected properly.

1.

Check the air circuit integrity and

 

 

reconnect.

 

2.

If the alarm doesn‘t get cleared, stop

 

 

treatment.

 

3.

Re-start treatment.

 

 

 

LCD: Circuit disconnected!

 

 

 

 

 

The air circuit is not connected properly.

1.

Check the air circuit integrity and

 

 

reconnect.

 

2.

If the alarm is not cleared, stop

 

 

treatment.

 

3.

Re-start treatment.

 

 

 

LCD: Low Minute Ventilation!

 

 

 

 

Minute ventilation level has dropped

Contact your clinician.

below the alarm setting level.

 

 

 

LCD: Low Respiratory Rate!, High Respiratory Rate!

 

 

The respiratory rate level has dropped

Contact your clinician.

below or has exceeded the alarm setting

 

 

level.

 

 

 

 

 

LCD: High Leak!

 

 

 

 

 

High mask leak for more than 20

Adjust the mask to minimise leak (see

seconds.

“Using mask-fit” on page 14).

 

Check the air circuit integrity and

 

reconnect.

 

If the problem persists contact your

 

 

clinician.

 

 

 

English

Troubleshooting 23

Image 27
Contents User Guide StellarRespiratory Care Solutions Making quality of care easy Contents Data management Cleaning and maintenance Adverse effects IntroductionIndications for use ContraindicationsStellar at a glance ResMed USB stick Internal batteryPatient interface HumidificationSetting up for noninvasive use Remote AlarmMobile use Attaching the H4i heated humidifier for noninvasive use Setting up for invasive useHmef Working with other optional accessories Attaching the pulse oximeterAdding supplemental oxygen Starting therapy using oxygen Stopping therapy using oxygenUsing the FiO2 monitoring sensor Attaching an antibacterial filter Connecting a sensorAbout the control panel Stellar basicsStarting therapy LCD screenPerforming a functional test Stopping therapy Starting therapyWorking with alarms Turning off the powerUsing mask-fit Tailoring treatment setup optionsUsing the menus Setup menu Options Setup menuSetup menu Clinical Settings Mask Type Setup menu Alarm Settings Alarm VolumeSetup menu Configuration Menu Parameter DescriptionOptions On, Auto Device Information Info menuEvent Summary Used HoursCleaning and maintenance Data managementDaily Weekly Monthly Replacing the air filter Multipatient use DisinfectionTroubleshooting Alarm troubleshootingServicing LCD Low Pressure Problem / possible cause Action LCD Blocked TubeLCD High Temperature 10, 11, 12 LCD High PressureLCD Apnoea Problem / possible cause Action LCD Non-Vented MaskLCD Internal Battery Low LCD Internal Battery UseOther troubleshooting USB stick is not readable or writeable Problem / possible cause SolutionDevice does not stop when you remove your mask Learn Circuit failedTechnical specifications Housing construction Power supplyExternal DC Power Supply isolated Internal BatteryEnglish Vrms Symbols Mask should never be blocked General warnings and cautionsGeneral warnings and cautions Product Warranty period Limited warrantyRespiratory Care Solutions Making quality of care easy EUR1