If pressure exceeds the scale range Out will appear on the display.

ZEROING

In order to achieve maximum accuracy it is recommended to zero the instrument in the orientation it will be used before taking any measurements.

Models 2000P, 2001P, 2002P, 2003P, 2020P, 2021P, 2022P, 2023P Ensure both ports are open to atmosphere. If indicated pressure is within 1%f.s (3% for Models 2000P, 2020P and 2080P) of the factory calibrated zero, press 0.0 key which will adjust the reading to zero.

Models 2080P, 2081P, 2082P, 2083P, 2084P

As above but press key instead of the 0.0 key.

ZEROING ABSOLUTE

This feature enables correction of the indicated ambient pressure between 900mbar and 1100mbar.

Models 2005P, 2025P

Open port to atmosphere and the instrument will indicate ambient pressure. Press 0.0 key and 900mbar (or the equivalent in other engineering units) will appear on the display. Press and hold this key. The display will scroll and when required atmospheric pressure is displayed (maximum 1100mbar) release key. The current pressure being measured will then be adjusted to the displayed reading and used for the rest of the session.

Note: This setting will be lost when the instrument is switched off.

Model 2085P

Press key instead of 0.0 key.

DISPLAY HOLD

Pressing key freezes the display. The symbol will appear on

the display. Press key again to return to normal display.

HOSE FITTING

The input port(s) on the instrument are designed to accept a hose with 4mm I.D. and 1mm wall. The hose is fitted by unscrewing the clamping nut, sliding the hose on to the stepped shaft and refitting the nut.