Note: Do not tighten fittings attached to the pressure ports to a torque greater than 2Nm otherwise the pressure seals may be damaged.
HAND PUMP OPERATION (if applicable)
Use the T piece to connect the hand pump to the instrument’s + port and to the equipment being tested. Exhaust the system by turning the release valve on the hand pump
Close the release valve and apply pump action to the hand pump piston to generate the required approximate pressure. Fine adjustment of the desired pressure can then be made by rotating the threaded piston. The release valve may be used to vent the system. Note: The hand pump is capable of generating 7bar max. It is, therefore, essential that extra care is taken not to
A pump must not be used on models 2000P, 2020P and 2080P
PLUS FUNCTIONS
Models 2020P, 2021P, 2022P, 2023P, 2024P, 2025P, 2080P, 2081P, 2082P, 2084P, 2085P
SMOOTHING MODE
Press = key and the instrument continuously averages the last four measurements to give a more stable reading on the display. Press = key again to return to normal measurement.
UNITS OF PRESSURE
Press key to select required units of pressure. Chosen units will be saved when the instrument is switched off.
Please refer to the table on the back page.
The 20x0P
This function will not operate while in store mode. Note: 2024P - 2084P
K = x 103 when using in H2O and Pa. M = x 106 when using in H2O and Pa.
MAX/MIN RECORDER
Press key to start. The instrument will display alternating symbols and current reading. Press again to display the
maximum temperature reading and symbol will be displayed. Press again to display the minimum temperature reading and symbol will be displayed. Press again to return to the actual temperature display.
Press X key to display the average temperature since the MAX/MIN
mode was activated and symbol will appear on display. Press
Xkey again to return to actual temperature display.
NOTE: To reset MAX/MIN recorder function switch the instrument off.
NOTE: Some functions may be inhibited while information is stored in either Auto, Manual or Max/Min modes. This is to prevent inadvertent parameter mixing whilst in any of the above modes. To