8494/95/96G/H Operating And Service Manual 19
Service
3. Allow the jaws of the tool used to close and pull straight back from the
connector without twisting. The contact should come with the tool. If
not, repeat the process. Do not reuse the contact.
4. Snap in a new contact by pushing it in place. Test the action of the new
contact by pushing it in. It should spring out again when released.
Type N Connector
Dimensions For critical dimensions for the type-N connector, see below.
Figure 7 Type-N Connector
Replacing the
Connector Outer Shell
NOTE The connector outer shell can be replaced only on the Option 001 (type-N
female) or the Option 003 (7 mm). The outer shell on the Option 002 (SMA)
cannot be replaced in the field. If these connectors are damaged, return the
instrument to Agilent Technologies for repair.
The connector outer shells on the 0ption 001 and 003 may be replaced as
follows:
1. With a 9/16-inch (1/2-inch for 7 mm) thin open-end wrench, unscrew the
outer connector body.
2. Replace the connector outer shell. See Table 5 on page 15 for
replaceable parts numbers.
3. Tighten the connector with the same wrench called out in step 1.