Operating and Programming

5071A Command References

5071A Command References

This section describes the additions, changes, and deletions to the information that appears in chapter 5 of the 5071A Operating and Programming manual.

5071A Command Summary

SCPI Commands

Table 5-2. SCPI Command Summary

Option 048: The response form for DIAGnostic:STATus:SUPPLY? command has changed:

AC/DC/LOW

Any TCO Option: The operation of [SOURce]:PTIMe:SYNC with the REAR parameter will always result in sync timed out with arming terminated after 1.5 seconds,

SCPI Subsystem Commands

The changes for the 5071A Command Summary above also apply to the corresponding SCPI Subsystem Command references,

Status Reporting

Table 5-5. SCPI OPERation Status Register

Option 048: In table 5-5. Bit 9 (On Battery) will always be zero (0).

Rear-Panel Status Output Operation

Status Output Programming

Option 048: The actual use of SCPI OPERation status register Bit 9 as a programmed condition will never cause a status output port state change.

The programming examples shown will work only when the 48 Vdc power option is absent, but are still valid to describe the technique for programming the status output port.

1-9

Page 21
Image 21