Assembly-Level Service
Service
2-7
2.1.1.1.2 A11/A11 Cable Test
Step 2 text is changed to:
Verify that the voltage present is between 45 and 51 Vdc with no
more than 1 Vrms ripple.
2.1.1.1.3 A11/A11-A1 Cables Test
Step 2 text is changed to:
Verify that the voltage present is between 45 and 51 Vdc with no
more than 1 Vrms ripple.
2.1.1.1.4 A1CR3 Bridge Rectifier Test
Step 2 text is changed to:
Verify that the voltage is at least 33 Vrms.
2.1.2.2.0 Ac-Line Fuse Fault Isolation: r/o LF1/T 1
The T1 part number w it hin If No response text to the QUESTION: is
changed to:
(9100-5134)
2.1.4.0.0 Ext. Dc Input Fuse Fa ul t
The second sentence of the first paragraph is changed to:
The assumption here is that ONLY the dc fuse blows, with the
instrument operating normally from the ac line.
2.1.5.0.0 No Operation From Internal Stand by Battery
This procedure is deleted.
2.1.5.1.0 Internal-Battery Charger Test
This procedure is deleted.
2.1.5.2.0 A11 F1 Fuse F a ul t
This procedure is deleted.
Option 048:
Option 048:
Option 048:
Option 048:
Option 048:
OPTION 048:
OPTION 048:
OPTION 048: